Micsig DP10013 Скачать руководство пользователя страница 2

Rise Time

Maximum Differential 

Test Voltage

DC+AC  PK-PK

Maximum input 
common mode voltage

Input referred noise

Common Mode 
Rejection Ratio

Input Impedance

Output Voltage

Overrange Alarm

Input cable length

Output cable length

Model

Bandwidth

Attenuation

Gain accurecy

Power Supply

Power

Dimension

160mVrms

200X

920 mVrms

2000X

50M

Ω

/1.25pF

25M

Ω

/2.5pF

single-ended 

to ground

differential

 DP20003

100MHz

3.5ns

200X

2000X

±

2%

560V (200X)

5600V (2000X)

80dB

DC

60dB

100KHz

50dB

1MHz

3V

Button light flashes

DC  5V

USB Supply

0.85W

14.5cm*6cm*2.7cm

Approx 45cm

Approx 90cm

0

~40

10%~85% RH

40mVrms

50X

230 mVrms

500X

10M

Ω

/1pF

differential

5M

Ω

/2pF

single-ended

 to ground

 DP10013

100MHz

3.5ns

50X

500X

±

2%

130V (50X)

1300V (500X)

CAT     1000V

80dB

DC

60dB

100KHz

50dB

1MHz

3V

Button light flashes

DC  5V

USB Supply

0.85W

14.5cm*6cm*2.7cm

Approx 45cm

Approx 90cm

0

~40

10%~85% RH

CAT       1000V

II

III 

Please read this manual carefully before use!

Introduction

The Micsig DP series of high voltage differential probes are designed for
the safe measurement of high voltage and floating voltage signals. Two 
models with 100MHz bandwidth are available, each capable of two different 
attenuation ranges: 500X(2000X) for a maximum test voltage of 1300V
(5600V) and 50X(200X) for amaximum test voltage of 130V(560V).

1

2

Specification

Operating  Humidity 

Operating 
Temperature

Panel

 

Description

Signal Input

Attenuation Range 
Buttons/Indicators
(Over-range if flashing) 

Signal Output  

Power Input 

Power Output 

Making Measurements

        Connect the USB input of the probe to the USB port of the oscilloscope 
       or a suitable USB power source.

       

4

3

1)    

Powering the probe:

2)    

Connecting the probe to the oscilloscope:

Connect the probe output BNC to the oscilloscope channel input.
Note: make sure the oscilloscope is properly grounded if necessary.

3)    

Set the appropriate attenuation range according to the measured 

       voltage.

 

4)    

Connecting the input to the device under test:

Using the appropriate input accessory, connect the probe to the device 
under test to start the measurement. If the attenuation range indicator 
flashes (indicating over-range), please immediately disconnect the 
power and input.

5)    

Set the measuring instrument.