Figure 4.19: Implant Settings
1. Select the button adjacent to
Therapist Trigger Mode
. This button will now toggle to
On
(see Figure 4.18)
.
Therapist Trigger Mode
must remain
On
to initiate a therapy session and trigger
stimulation per the
VNS Pairing Therapy
menu (see
Section 1.1.1:
).
2. Optionally, you may change the stimulation parameters displayed in the
Implant Settings
menu (see Figure 4.19). See
3. After all desired parameters have been set, press the
Download Settings to IPG
button
to download the chosen parameters to the implant.
Initiation of Paired VNS
™
using
Therapist Trigger Mode
is detailed in
Section 5:
4.
It is recommended to verify that the VNS settings chosen are tolerable for the patient. See
4.5
Impedance
In the
Impedance
portion of the
Patient Information
area of the
Program Implant
menu, the
Lead impedance can be checked by selecting the
Check
or
Lead Impedance
buttons (circled
in Figure 4.20). Doing so delivers a small current through the Lead to calculate the Lead
impedance. This measurement pulse contains minimal charge and is not expected to cause any
discomfort, even in the most sensitive patient. The
Lead Impedance
button is available at all
times once the user is logged in. See
Содержание 2100
Страница 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Страница 28: ...26 0002 0005 0 en 2022 28 Figure 4 22 Low Lead Impedance Notification Figure 4 23 Lead Impedance Displayed ...
Страница 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Страница 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Страница 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...