4 P
ROGRAMMING THE
I
MPLANT
The
Program Implant
menu (see Figure 4.1) allows the user to set the patient information, test
and set stimulation parameters, set therapy mode(s), and check impedance during implant and
follow-up visits.
Figure 4.1: Program Implant
4.1
Connecting to a Specific IPG
Perform these six steps before therapy for each patient to connect to their specific IPG.
1. Select the
Disconnect
button to disconnect any IPG that is connected to SAPS that is
different from your patient’s IPG Serial # (see Figure 4.2). Select “Continue” when
prompted to confirm.
Figure 4.2: Disconnect
2. Verify there is no Serial # displayed in SAPS; you will see 3 dashes next to “Serial” if this
is the case (see Figure 4.3).
Figure 4.3: Blank Serial #