MICROTEST
22
S1
S3
C/-:
Press
under Insulation Test Setting to enter High voltage Channel
Setting text box. Use navigation key to move the cursor to the channel
and press
confirm.
to start the setting. After the setting, press
to
Figure 5.5.3.2
Voltage / Frequency:
Move the cursor to Voltage and use numeric key to set up
voltage value, the maximum value is AC 5.0kV, DC 6.0kV, the
minimum voltage is 0.1kV. After the setting, press navigation
key to move the cursor to frequency, use S1 to select frequency
Dwell:
High voltage output sequential times. Use numeric key to set up, maximum
value is 99.9 second, minimum value is 0.1 second
Ramp:
The maximum value is 99.9 second, the minimum value is 0.1 second
Max/Min:
Use numeric key to set up values, the maximum value for AC is 10mA, DC
is 5mA, the minimum value is 0 for both AC and DC
Arc Sen.:
The smaller value, the higher sensitivity. Use numeric key to set up
values, if value is 0 then the function is off
Offset:
To deduce the leakage voltage which made by testing equipment or fixtures.
Use numeric key to enter the leakage voltage value and tester will deduce the
error automatically
DUT No.:
Use numeric key to set up, the maximum value is 4, the minimum value
is 0. Use 0 at the first step, after the first step the minimum value should
S1