7  Using the Scanner ICC Profiler Program

1. Introduction

The Scanner ICC Profiler is a scanner calibration and
profiling utility program designed exclusively for use
with Microtek scanners, including the ArtixScan 1100.
The Scanner ICC Profiler lets you calibrate the color
attributes of your scanner and lets you create an ICC
color profile customized and tailored especially for your
Microtek and Artix scanner.

Why do you need the Scanner ICC Profiler?

For each Microtek and Artix scanner, a generic factory-set
ICC color profile is built in and included for use with the
scanner controller program — in this case, ScanWizard
Pro. This factory-set profile delivers high color quality in
general. But scanners, like any high-end device, interpret
colors differently, resulting in subtle variations and
differences in color imaging. In addition, individual
scanners may also perform different due to other factors,
such as aging of the scanner lamp or ambient temperature.

With the Scanner ICC Profiler, such color variations and
differences can be minimized or eliminated, as the
program calibrates the scanner and compensates for subtle
color deviations. The customized color profile resulting
from the use of the Scanner ICC Profiler, in effect,
delivers more accurate color and contributes to overall
improved image quality.

To keep the colors in your scanner consistent over time,
it’s recommended that you use the Scanner ICC Profiler
regularly. Professional photographers or others who
require extremely precise color may wish to use the
Profiler every time before they use their scanner, but for
most general usage, calibrating twice a month should be
sufficient.

2. Calibration target

Your Scanner ICC Profiler kit includes two industry-
standard IT8 color targets for calibration.

•  Kodak Q-60R1: This is a reflective target for calibrating

the upper scan bed. Calibrating the upper scan bed
ensures color accuracy when you scan reflective
materials such as photos and prints.

•  Kodak Q-60E1: This is a transparent target for

calibrating the lower scan bed. Calibrating the lower
scan bed ensures color accuracy when you scan
transparent media such as slides and filmstrips.

3. Taking care of the target

The calibration target is very delicate and must be handled
carefully. Make sure you follow these rules in caring for
the target.

•  Gently take the target out of its protective sleeve.

•  Do not touch the target image with your fingers or with

any other object.

•  When not in use, keep the target out of light — even

interior lighting.

•  Always return the target to its protective sleeve

immediately after use.

•  Store the target away from light in a cool, dry place,

since long exposure to heat and light can change the
colors on the target.

4. Replacing the target

With proper care, your target should hold its original
colors for about a year or so starting from the date of use.
After that, it will begin to fade, as is the nature of most
photographic materials. The change may be unnoticeable
to the naked eye, but even very slight changes in the target
decrease the performance of the Scanner ICC Profiler. To
maintain color accuracy, Artix recommends that you
replace the target once a year, or when you notice a loss
of image quality in your scanned images.

Содержание Artix Scan1100

Страница 1: ...English ArtixScan1100 HardwareUser sGuide...

Страница 2: ...ernational Inc 6 Industry East Road 3 Science Based Industrial Park Hsinchu 30077 Taiwan R O C Tel 886 3 5772155 Fax 886 3 5772598 Internet http www microtek com Microtek Europe BV Max Euwelaan 68 NL...

Страница 3: ...Operating Environment 6 3 InstallationontheMacintosh 7 A Installation for Macintosh computers with built in SCSI ports 7 B Installation for the new G3 Macintosh computers 8 C Installing the software 8...

Страница 4: ...tosh 26 B For Windows 98 95 NT 4 0 26 6 Placing the color target 26 7 Calibration 27 8 Scanning the target image 27 9 Aligning the target registration marks 28 A Aligning the upper left registration m...

Страница 5: ...which you operate If you are operating under the Macintosh environment go to the General Installation section of the manual then proceed to Installation on the Macintosh then go to Operating the Scan...

Страница 6: ...uide English A Horizontal ruler Scanner cover Glass surface Vertical ruler Transparency tray ArtixScan 1100 Scanner Power switch Power LED green Transparency Ready indicator orange Reflective Ready in...

Страница 7: ...ArtixScan 1100 Hardware User s Guide 3 English SCSI 50 pin connector SCSI 25 pin connector ID switch Terminator switch ArtixScan 1100 Scanner ON OFF 6 Power connector...

Страница 8: ...a reasonable amount of abuse Refer to the package contents checklist included with your scanner to ensure that you received all of the parts necessary for scanner setup Should you observe any damage o...

Страница 9: ...scanner has not been damaged during shipment however you will need to perform the following preliminary power up test Contact your dealer immediately if the scanner is not in proper working condition...

Страница 10: ...f the power switches and disconnect the cord Beforeyouchoosealocationforthescanner pleasereadtheoperatingen vironmentrequirementslisted b e l o w Requirements for the Operating Environment Make sure t...

Страница 11: ...on for details on termination 3 Plug the power cord to the power connector at the back panel of the scanner and plug the other end of the power cord to your AC power source or wall outlet 4 Turn on yo...

Страница 12: ...it is seated all the way in the slot This is important as an improper card connection will make you unable to use your scanner and you will then have to remove the computer case and reinsert the card...

Страница 13: ...3 Termination A terminator is included in your scanner package and should be installed on your scanner in the following situations If the scanner is the only SCSI device in your SCSI chain or If you...

Страница 14: ...mputer then plug the power cord back in 5 Turn your PC back on 2 Checking the interface card status When Windows starts with the Adaptec SCSI card installed in your computer the card is detected and t...

Страница 15: ...Pro make sure you choose Adaptec SCSI Interface Card as your interface type After all the software has been installed shut down your computer 4 Connecting the scanner 1 Turn off your PC 2 Plug the 25...

Страница 16: ...k on the System icon in Control Panel and select Device Manager from the top 3 Double click on SCSI controllers to display the dialog box below 4 Click on the Adaptec AIC 7850 PCI SCSI Control ler opt...

Страница 17: ...flicting device list For example to resolve the Memory Range conflict take these steps a At the Resource type double click on the Memory Range string the Edit Memory Range dialog box is displayed b Wh...

Страница 18: ...Interrupt Request Memory Range and Input Output address settings If you are asked to shut down your computer select No then click Close You will be asked whether you wish to restart your computer Cli...

Страница 19: ...p 10 Double click on SCSI controllers to check whether conflict happens If a conflict exists follow the steps in Resolving Situation B 11 When the settings are all correct click OK to save the modific...

Страница 20: ...s that under Windows 98 95 2 Installing the Windows NT driver a Start your computer and log in as Administra tor Note You must be logged in as Administra tor in order to install the scanner software c...

Страница 21: ...ivers tab and click on the Add button to select Adaptec on the left and Adaptec AIC 7800 PCI SCSI Controller on the right 5 Click OK to bring up the following screen The Adaptec card should now be lis...

Страница 22: ...when you use the upper scan bed of the ArtixScan 1100 The upper scan bed is used for scanning reflective materials such as photos and prints A General scanning 1 Lift the scanner lid 2 Place the image...

Страница 23: ...your scanner package See below for more details Using the Universal Glass Film Holder Use the Universal Glass Film Holder together with the transparency tray when you wish to scan non standard sized...

Страница 24: ...nsparent film place the film face down on top of the glass surface of the Universal Glass Film Holder 2 To secure the film to the Universal Glass Holder place the vinyl strip on the edge of the transp...

Страница 25: ...ing the 35mm Slide Holder 1 Insert the individual 35mm slides to be scanned into the 35mm Slide Holder 2 Place the 35mm Slide Holder in the Main Holder 3 Insert the whole assembly into the transparenc...

Страница 26: ...nsparency bay the drawer or lower compartment of the scanner C Using the 35mm Filmstrip Holder 1 Insert the 35mm filmstrip to be scanned into the 35mm Filmstrip Holder a Push to open the 5 piece 35mm...

Страница 27: ...he 6 x 9 cm Film Holder a Push on the side to open the holder b Place the film in the holder c Pull the side down to close the holder 2 Place the 6 x 9 cm Film Holder in the Main Holder 3 Insert the w...

Страница 28: ...the 4 x 5 Film Holder a Push on the side to open the holder b Place the film in the holder c Pull down the side to close the holder 2 Place the 4 x5 film holder in the Main Holder 3 Insert the whole a...

Страница 29: ...remely precise color may wish to use the Profiler every time before they use their scanner but for most general usage calibrating twice a month should be sufficient 2 Calibrationtarget Your Scanner IC...

Страница 30: ...ck on the Install option for the Scanner ICC Profiler program then follow the on screen instructions until the program finishes installing 6 Placingthecolortarget Positioning the target in the upper b...

Страница 31: ...w a message appears informing you of the next step to be done To select the target image move the pointer now a crossbar to the preview image and draw a frame enclosing the entire target image When yo...

Страница 32: ...the upper left mark is aligned the upper right part of the target image is displayed and an instruction dialog box prompts you to align the upper right registration mark Upper left registration mark...

Страница 33: ...you the results If the process was successful a custom scanner ICC profile is saved automatically to your system s ColorSync Profiles folder Exit the Scanner ICC Profiler now and use the new scanner p...

Страница 34: ...mand lets you choose the correct ScanWizard Pro Plug in driver for the ICC Profiler program in case the Plug in driver cannot be found b Get Current Scanner Info This command provides information on t...

Страница 35: ...ow your scanner sees color 4 Tonal Mapping This option lets you select the way in which the tonal reproduction curve of the profile is controlled Normal Slightly brightens the highlights but also dark...

Страница 36: ...d the required alignment of the three registration marks on the target 2 Previous Mark This button lets you go back one step to align the previous registration mark 3 Next Mark This button lets you go...

Страница 37: ...o the colori metric data inside the TDF to create an ICC color profile a characterization of how your scanner sees color 4 Tonal Mapping This option lets you select the way in which the tonal reproduc...

Страница 38: ...ner 2 Scanner Probe This command displays the SCSI devices on the SCSI chain as well as the SCSI ID numbers of the individual devices To use this command 1 Choose the Get SCSI Chain Info command 2 If...

Страница 39: ...d the required alignment of the three registration marks on the target 2 Previous Mark This button lets you go back one step to align the previous registration mark 3 Next Mark This button lets you go...

Отзывы: