background image

bolt on the top of plug with little screwdriver and pull out the plug 

from transmitter.

 

b)  To disconnect the plug, please take out the central bolt on the top 

of  plug,  then  turn  the  plug  to  the  bottom  and  insert  a  little  flat 

screwdriver into a corner signed “Lift” and give a force. The plug 

core will be disconnected with housing. 

c)  To  connect  the  cable,  put  the  cable  through  the  cable  jack  and 

connect  the  wires  with  terminals  on  the  plug  core  correctly  (the 

connection  terminals  are  signed  with  clear  numbers).  Please 

choose  Φ4.9mm  shield  cable  and  connect  the  cable  reliably  to 

prevent short circuit. 

d)  Pull  the  cable  slightly  and  push  the  plug  core  into  housing  (a 

rattling sound could be listened), then screw down the cable-fixed 

nut.   

e)  To remove the cable, loose the cable-fixed nut to relax the cable, 

and operate as Item2 to disconnect the plug. Take out the cable 

from  terminals  with  little screwdriver  and  pull  out the cable from 

cable-fixed  nut,  then  renew  the  connection  between  plug  core, 

plug and socket. 

Caution: renew the rectangle-ring to the pre-mounting situation in order 
tomaintain the protection class when connect plug and socket. 

5 Unpacking

Components and Storage 

Содержание MPM4841

Страница 1: ...MPM4841 Operation Manual...

Страница 2: ...tion manual due to renovation of production technology and craftwork If some information is changed no more notice will be edited Please pay attention to the latest version Our company also reserves t...

Страница 3: ...omatically tested by computer and compensated zero and temperature in wide range by laser trimming system The digital circuit processes the sensor signal into standard output signal The product is sui...

Страница 4: ...Signal Output 4mA 20mADC 2 wire 0 1V 5 10VDC 3 wire 0 5V 4 5VDC 3 wire Operation Temperature 30 80 4 pin Plug M12x1 Plug 20 70 cable material PE PVC 20 80 cable material PUR Storage Temperature 40 12...

Страница 5: ...3 B1 SW19 50 78 27 SW27 G1 8 30 SW19 27 SW27 30 G1 8 50 105 B2 G1 8 27 17 5 60 5 SW27 SW27 27 G1 8 17 5 90 5...

Страница 6: ...be mounted vertically up to the horizontal direction If the condition is unavailable the allowed max mount slope angle from transmitter to the horizontal direction is 30 degree It is not recommended...

Страница 7: ...hing hard and do not press the diaphragm with finger or something sharp to protect the diaphragm The recommended method is to see Fig 2 Pressure Transmitter Valve Pipe Fig 2 4 Electric Connection 4 1...

Страница 8: ...be blocked or water infl ow and it must be connected to the atmosphere to ensure the pr otection of the product and the accuracy of the output measureme nt The electric definitions of M12x1 Plug are a...

Страница 9: ...ed with clear numbers Please choose 4 9mm shield cable and connect the cable reliably to prevent short circuit d Pull the cable slightly and push the plug core into housing a rattling sound could be l...

Страница 10: ...cial Plug 1 Production Manual 1 Production Qualification Certificate 1 5 3 Storage The transmitter should be stored in dry ventilate room ambient temperature 40 120 20 85 Cable type and the relative h...

Страница 11: ...ot connect it at will to protect the transmitter c The transmitter and the external circuit connect through the speci al plug seat When the installation of electrical connection is comp leted be sure...

Страница 12: ......

Отзывы: