LVS-95XX Series Barcode Quality Station Operations Manual
LVS-95XX Series Barcode Quality Station Operations Manual - English
Page 44 of 94
Calibrated Conformance Standard Test Card Preventive
Maintenance Instructions
Place the Calibration Test Card in a clean place and away from direct sunlight. If the test card is nicked,
soiled, or damaged in any way, it must be replaced. Contact Microscan or your Microscan representative to
obtain a new Calibration Test Card.
Calibration Test Cards are valid for two years from the date entered in the “Date issued” field on the Test
Card. Valid test cards are required to obtain accurate grading results.
Each calibration card has been tested by a NIST traceable test device and the values for Decodability,
Contrast, Modulation and Rmax are recorded on the card.
Enter Test Card Parameters for a Replacement Card
•
Click
the
Enter test card parameters
button on the
Calibration Tab
screen. Click the
OK
button when
a warning box appears.
•
Enter the values for Decodability, Contrast, Modulation and Rmax in the gold fields located under the
column entitled “Actual.”
•
Place the new Test Card into the field of view and push the
Calibrate
button. If calibration fails, push the
Calibrate
button again. If calibration fails repeatedly, call Microscan.