XYZ CHANNEL RECORDING (7 ELECTRODES)
ELECTRODES
This electrode configuration is suggested for:
― Derived 12 channel ambulatory recordings*.
― High Resolution 20min recordings for late potential analysis; for this application a
shielded cable is recommended to reduce noise level.
White
X-
Fourth intercostal space in right midaxillary line
Red
X+
Fourth intercostal space in left midaxillary line
Black
Y-
Superior aspect of the manubrium
Brown
Y+
Lower ribs
Blue
Z-
Directly posterior to V2
Orange
Z+
Standard V2 position
Green
ref
Over 8th rib in right midclavicular line
Cable length: 152 cm
* Available with SyneScope Holter software.
ELECTRODE PLACEMENT
For an ambulatory orthogonal ECG recording close to the XYZ late potential configuration,
the following hook-up is recommended. However, due to constraints of daily activities, some
minor variations may be introduced, provided that polarity of electrodes and axis orientations
are maintained.
Y-
Y+
X+
ref
Z+
Z-
X-
TWELVE-LEAD RECORDING* (10 ELECTRODES)
ELECTRODES
1
Yellow
LA
Left Anterior acromium
2
Red
RA
Right Anterior acromium
3
Green
LL
Left Interior iliac crest
10
Black
GND
Right Interior iliac crest
4
Purple
V1
V1 position
8.4.
8.4.1.
8.4.2.
8.5.
8.5.1.
8. POSITIONING THE ELECTRODES
18
SpiderView – UA10709B