12. Troubleshooting
When you receive an error code, please retest. If you receive a
second error code, then contact your local distributor and test the
patient with a laboratory method. Do not interpret an error code
as a patient result in any case.
Error Code
Description
Corrective Actions
E001
Low battery.
Please use power adapter
or the base as a power
source.
E001.1
Battery failure.
Please turn off the meter
and restart it. If the
problem recurs, please
contact the relevant
technicians.
E002
Heating has timed out.
Please adjust the ambient
temperature to 10
℃
-
35
℃ ,
and turn off the
meter for 5 minutes, then
turn on the meter and
restart the test. If the
problem recurs, please
contact the relevant
technical staff.
E003
The ambient temperature
is too high or too low.
Please ensure that the
meter is used in an
environment of 10
℃
-
35
℃ .
E004
The added sample is
insufficient.
Please retest with a new
test strip and ensure
that the added blood is
sufficient for the test.
E005
Internal QC failed.
Please use an unexpired
test strip and enter
the correct test strip
Codechip number.