ZERO SETTING

The following instructions allow you to set 
the zero on the instrument:

1) Switch the level on and select the lower 
scale (A). Then with the potentiometer on 
the right, look for the zero (the value is 
shown on the digital display also). Then 
verify the zero on the higher scale (E).

2) Turn the level of 180°, read the value 
on the display and adjust the potentiome-
ter untill when you reach the half of the 
read value. (Example: if you read 26 µm 
in the position (a), adjust the potentiome-
ter to 13 µm on position (b)). 

3) Repeat the same steps (1+2) by turning 
the level of 90°, as shown hereunder.

The instrument is now ready for use.

      
         

a                    b                  c           d

In the pictures the 4 step for zero setting

Measure scale 

selector 

Potenziometer

(zero adjustment)

MEASURE SCALES

The instrument allows you to work with 
five different measure scales, as follows:

Scale A

 = 1 div 250 µm/m

Scale B

 = 1 div 50 µm/m

Scale C

 = 1 div 10 µm/m

Scale D

 = 1 div 5 µm/m

Scale E

 = 1 div 1 µm/m

div = division

Содержание LE051

Страница 1: ...LE051 ELECTRONIC LEVEL USER S MANUAL your partner for metrology...

Страница 2: ...2 2016 Microplan Documento LE051_IT Versione 16 04 All rights reserved Microplan 2016...

Страница 3: ...m always hangs ver tically thus allowing measurement of the displacement relative to the horizontal base by means of an inductive position transducer The innovation that applies to all MICRO PLAN leve...

Страница 4: ...data transmission to PC RoHS compliant Lagammecompl tedesniveaux lectroniquesMicroplan Unit de lecture num rique analogique ou analogique num rique Syst me innovant de m canique dans bain d huile pou...

Страница 5: ...tic fields as the measurements could be altered The surface plate to be checked should be levelled as closely as possible to avoid inaccuracies of the readings due to com pound errors Check the level...

Страница 6: ...sitivity is as high as 1 m m 0 2 second of arc The internal software manages the whole acquisition process the measures avera ging and the data display on a 16 charac ters LCD and a galvanometer for a...

Страница 7: ...ccessories LE051 Art LE0433 LE051 Unit Battery charger included with unit LE0433 Art LE0435 Serial connection cable between level and PC optional Art LE0439 WPLAN Microplan Flatness Software opzional...

Страница 8: ...ter untill when you reach the half of the read value Example if you read 26 m in the position a adjust the potentiome ter to 13 m on position b 3 Repeatthesamesteps 1 2 byturning the level of 90 as sh...

Страница 9: ...advise customer against using other battery chargers which could damage the inner batteries of the level The battery charger state is shown on the digital display when the level is switched on with th...

Страница 10: ...2 The cable connects the LE051 unit to the pc through the serial gate COM1 o COM2 and needs the use of WPLAN MICROPLAN Flatnesssoftwa re release 4 I Regarding the complete procedure for the memorized...

Страница 11: ...019 VARALLO Vercelli ITALY 39 0163 54619 569084 39 0163 564081 mpi microplan group com Skype microplan italia www microplan group com TECHNICAL SUPPORT By doubt or information regarding the working of...

Страница 12: ...54619 569084 FAX 39 0163 564081 EMAIL mpi microplan group com SKYPE microplan italia MICROPLAN FRANCE SAS 2 LA GRAULE F 23360 LA FORET DU TEMPLE FRANCE TEL 33 5 55806666 FAX 33 5 55806620 EMAIL mpf mi...

Отзывы: