4.

 

Installing the MPA-20 

 
4.1

 

MPA-20 12 mm Installation 

 

The  MPA-20  can  be  installed  to  motorize  manual  drives,  mirror  mounts  or  other 
applications.  While  installing,  support  the  MPA-20  by  holding  the  motor/gear 
reduction unit to avoid applying torque to the housing. In place of the spherical 
actuator tip, a 4-40 thread may be used for pulling applications. When preparing 
the  stage  for  vacuum  environments,  take  the  necessary  precautions  (such  as 
wearing  gloves,  clean  room  and  clothing,  etc.)  when  handling  the  stage  as  to 
avoid any contaminants. 

 

 

 

 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Requires: 
1x MPA-20 Linear Actuator 
1x Locking Nut 
1x 17mm Hex Wrench (not supplied) 
 
 

1.  

Unscrew Locking 
Nut from MPA-20 

2.  

Insert MPA-20 into 
Mounting Surface 

Customer Supplied 
Mounting Surface 

3. 

Tighten Locking Nut 
using 17mm wrench 

Locking Nut 

Содержание MPA-20 series

Страница 1: ...S e r i e s Reference Manual Micro Precision Actuator UHV MPA 20...

Страница 2: ...MPA 20 Linear Actuator Reference Manual Rev 1 01 MICRONIX USA LLC 15375 Barranca Parkway E 106 Irvine CA 92618 Tel 949 480 0538 Fax 949 480 0538 Email info micronixusa com http micronixusa com...

Страница 3: ...bers 4 3 Preparing to Install the MPA 20 5 3 1 Installation Preparation 5 3 2 Package Contents 5 4 Installing the MPA 20 6 4 1 MPA 20 12 mm Installation 6 5 Technical Specifications 7 5 1 Electrical C...

Страница 4: ...manual drive stages and mirror mounts It utilizes a 2 phase stepper motor and two mechanical limit switches The MPA 20 is compatible with the MMC 200 controller Features Travel range of 12 mm Gear red...

Страница 5: ...Technical Data Motor SM 005 Screw Pitch mm 0 5 Gear Ratio 4 1 Speed Max mm sec 1 25 Resolution Calculated m 0 625 with micro stepping Resolution Typical m 0 1 Bi directional Repeatability m 1 Uni dir...

Страница 6: ...MICRONIX USA LLC Irvine California www micronixsusa com 2 Model Configurations 2 1 MPA 20 Order Numbers Contact MICRONIX USA for custom versions Order No MPA 20 1 1 0 1 9 SM 005 1 12 mm 1 No Encoder...

Страница 7: ...face which is free of debris burrs or dings An indoor atmosphere free of corrosive gases excessive dust and condensation Temperature range of 0 40 C Relative humidity between 20 80 Locate away from wa...

Страница 8: ...applying torque to the housing In place of the spherical actuator tip a 4 40 thread may be used for pulling applications When preparing the stage for vacuum environments take the necessary precautions...

Страница 9: ...itch Common 6 Limit Switch near motor 7 Limit Switch 8 Thermal Element 9 Thermal Element For compatibility with MICRONIX USA controllers Pin 5 7 and shield are internally connected Negative limit only...

Страница 10: ...PA 20 Linear Actuator The mechanical limit switches are factory calibrated for a minimum travel of 12mm when the switches are initially activated Customer adjustment of the limit switches is not recom...

Страница 11: ...MPA 20 Micro Precision Actuator Rev 1 01 9 Reference Manual MICRONIX USA LLC Irvine California www micronixsusa com 5 3 Dimensions 5 3 1 MPA 20 12 mm...

Страница 12: ...he actuator is in motion to avoid personal injury Follow the Safe Operating Environment requirements and use proper cable management to ensure a clean and safe operating environment The MPA 20 is desi...

Отзывы: