2 Smart VacPrep Hardware Installation
C
ONNECT
E
THERNET
C
ABLE
C
ONNECT
O
NE
A
NALYZER
1. Plug one end of the shielded Ethernet cable into the Ethernet port on the analyzer.
2. Plug the other end of the shielded Ethernet cable into the Ethernet port on the computer.
Connectivity for the 067 Smart VacPrep is different. The 067 Smart VacPrep must be
connected using an Ethernet switch. To connect the 067 Smart VacPrep to the
analyzer, follow the instructions under
Connect Multiple Analyzers with an
.
C
ONNECT
M
ULTIPLE
A
NALYZERS WITH AN
E
THERNET
S
WITCH
An Ethernet switch is required. The Smart VacPrep unit will be attached to the
Ethernet switch. The Ethernet switch will then be connected to the computer using a
shielded Ethernet cable.
An Ethernet switch is required to connect:
n
Multiple analyzers
n
One or more analyzers and one computer
To establish the connection:
1. Connect the power cord of the Ethernet switch to an appropriate power outlet.
2. For each analyzer, connect a shielded Ethernet cable from the Ethernet port of each ana-
lyzer to a numbered port on the Ethernet switch taking care not to use the uplink port on the
switch.
3. Use a shielded Ethernet cable to connect the computer to the Ethernet switch. Do not use
the uplink port.
4. Ensure the computer and the Ethernet switch are powered on.
2 - 2
Smart VacPrep Installation Instructions and Checklist
067-42870-02 (Rev A) — June 2022