Gemini VII
Introduction
239-42828-01 - Dec 2012
3-1
3. OPERATING PROCEDURES
Introduction
This chapter contains information on:
• specifying sample defaults
• creating sample files in Advanced, Basic and Restricted formats
• preparing samples and performing an analysis
• printing file contents, listing file statistics, exporting isotherm data, and converting sample files
• generating graph overlays
Specifying Sample Defaults
Sample information files include the information required by the analyzer to perform analyses and col-
lect data. When a sample information file is created, the system assigns the default values specified in
either Advanced or Basic format.
Format
Description
Advanced
Displays all parts of sample information and parameter files. Navigate to
parameter screens by clicking the tabs across the top of the window.
Basic
Displays the entire sample information file in a single window with no tabs
across the top. This option is used once the parameter files have been created.
The previously entered or default parameter files are then accessible using
dropdown lists.
Restricted
Displays the sample information file in a single window similar to the Basic
format with certain functions disabled. The Restricted format requires a
password when entering or exiting the Restricted format. It is typically used in
laboratories where analysis conditions must remain constant, for example, in the
pharmaceutical industry. The Advanced format is not available from the
Restricted format.
Содержание Gemini VII 2390a
Страница 1: ...Gemini VII Operator s Manual V3 02 239 42828 01 Jan 2014...
Страница 4: ......
Страница 12: ...Table of Contents Gemini VII viii 239 42828 01 Nov 2013...
Страница 36: ...Menu Structure Gemini VII 2 14 239 42828 01 Dec 2012...
Страница 192: ...Service Test Gemini VII 6 24 239 42828 01 Dec 2012...
Страница 217: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 25 Isotherm Linear Plot...
Страница 218: ...Report Examples Gemini VII 7 26 239 42828 01 Dec 2012 BET Surface Area Report...
Страница 219: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 27 BET Surface Area Plot...
Страница 220: ...Report Examples Gemini VII 7 28 239 42828 01 Dec 2012 t Plot...
Страница 221: ...Gemini VII Report Examples 239 42828 01 Dec 2012 7 29 BJH Adsorption Cumulative Pore Volume...
Страница 222: ...Report Examples Gemini VII 7 30 239 42828 01 Dec 2012 BJH Adsorption dV dlog w Pore Volume...
Страница 264: ......
Страница 266: ......
Страница 294: ...Appendix B Gemini VII B 28 239 42828 01 Nov 2013...
Страница 364: ...Appendix F Gemini VII F 18 239 42828 01 Dec 2012...
Страница 372: ...Gemini VII Smart VacPrep Operator Manual Dec 2013 G 1 G Smart VacPrep Operator Manual...
Страница 373: ...OPERATOR MANUAL Smart VacPrep Operator Manual Nov 2013 SMART VACPREP INTELLIGENT VACUUM SAMPLE PREPARATION SYSTEM...
Страница 380: ...Contact Us vi Smart VacPrep Operator Manual...
Страница 390: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Страница 424: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...