Sample Preparation
Feb 2013
2-19
Alternating Vacuum-Backfill Preparation
Alternating vacuum-backfill sample preparation utilizes a combination of vacuum and gas
backfilling techniques to remove contaminants and prepare samples for analysis. The VacPrep
061 is designed for low-pressure backfilling only.
1. Perform steps 1 through 8 as described in the previous topic,
.
2. Allow the sample to degas for the desired length of time, for example, 10 minutes.
3. Leave the sample tube in the heating station and turn the VAC/ GAS control knob to the
GAS position.
4. Allow gas to flow into the sample tube for 5 to 10 seconds.
5. If you are preparing other samples on the degasser which are not to be evacuated, turn the
corresponding VAC/GAS control knobs to the OFF position.
6. Turn the VAC/GAS control knob for the sample being prepared to the VAC position.
7. When the vacuum reading falls below 100 millitorr (100
mHg), turn the VAC/GAS
control knobs for any other samples to be evacuated back to VAC.
8. Repeat steps 2 through 7 until the sample is sufficiently degassed.
Do not exceed 5 psi regulator pressure when backfilling sample tubes. If the gas
pressure is greater than the recommended maximum of 5 psig (35 kPag), the sample
tube could be ejected from the fitting or broken.
Some fine-powdered samples such as controlled pore glass, fluid-cracking catalysts,
and carbon may be difficult to prepare by vacuum preparation. Such materials tend to
fluidize and float up the sample tube toward the vacuum source. Micromeritics
recommends that such samples be prepared by flowing gas preparation and, if
necessary, then subjected to vacuum preparation.
If sample has been accidentally sucked into the instrument, follow the instructions for
cleaning and replacing vacuum fittings in Chapter 3.
The vacuum fitting for the sample tube may be hot enough to cause burns after the
sample has been heated. Wear the cotton gloves supplied in the accessories kit and
handle the fitting carefully to avoid burns.
Содержание 061-00023-00
Страница 1: ...VacPrep 061 Operator s Manual 061 42802 01 Feb 2013 ...
Страница 2: ... Micromeritics Instrument Corporation 2002 2013 All rights reserved ...
Страница 4: ......
Страница 30: ...Sample Preparation VacPrep 061 2 10 Feb 2013 TriStar 3Flex 1 4 3 8 1 2 in 9mm 12mm Stem P N 065 25815 00 ...
Страница 46: ...Sample Preparation VacPrep 061 2 26 Feb 2013 ...
Страница 56: ......
Страница 60: ...Ordering Informatioin VacPrep 061 4 4 Feb 2013 ...