List of Recommended Reagents
Chemicals from A to Z
The following tables enlist the most frequently used reagents.
They provide useful information for the safe and efficient use of
MICROLIT BEATUS. All the safety precautions and recommendations
within this Operation Manual must be followed carefully.
Code explanations:
A = Good resistance
B = Acceptable with limitations
C = Not recommended
1 = Acid vapours (better resistance with lower concentration).
Do not leave the instrument on the bottle.
Rinse with distilled water after use.
2 = Risk of damage, softening or discoloration of external parts through vapours.
Do not leave the instrument on the bottle.
Rinse with distilled water after use.
3 = Chemical degradation of glass parts (plunger/barrel)
6.
Operating Exclusions
Do not use the instrument with:
Liquids that attack FEP, PFA and PTFE (e.g. dissolved Sodium Azide*)
Liquids that attack borosilicate glass (e.g. Hydrofluoric acid)
Hydrochloric acid > 40% and Nitric acid >70% | Tetrahydrofuran |
Trifluoroacetic acid
Explosive liquids (e.g. Carbon Disulfide)
Suspensions (e.g. of charcoal) as the solid particles may clog or
damage the instrument
Liquids that attack PP (cap)
* The permissible concentration of dissolved Sodium Azide is not more than 0.1%.
4.
•
•
•
•
•
•
Storage Conditions
Store the instrument and accessories in a clean, cool and dry place.
Storage temperature may vary from – 20°C to +50°C (from – 4°F to 122°F)
5.
Operating limitations
Liquids, which form deposits, may make it difficult for the piston to move or
jam it (like crystallizing solutions or concentrated alkaline solutions).
If the piston becomes too difficult to move, immediately clean the instrument.
(Refer to section 13)
While dispensing inflammable fluids, avoid the built up of static charge.
Make sure that you do not dispense into plastic vessels and do not wipe
instruments with a dry cloth.
MICROLIT BEATUS is designed for general laboratory applications and
complies with relevant standards, like DIN EN ISO 8655. Please check the
compatibility of the instrument for a specific application (e.g. trace material
analysis, food sector etc.). Approvals for specific applications, like production
and administration of food, pharmaceuticals and cosmetics are not available.
•
•
•