33
Do not iron the cuff!
Accuracy test
We recommend the WatchBP O3 device be tested for
accuracy every 2 years or after mechanical impact
(e.g. being dropped). Please contact Microlife to
arrange for an accuracy test.
Cleaning the cuff and straps
Use a mild disinfectant solution to clean the pouch,
straps, sling and belt. To wash these items and cuff
in washing machine, use a cuff plug (article #90-
273MZ11-004) to keep the inside of the bladder dry.
Fold and place the cuff inside a washing bag. Wash
these items using warm water and a mild detergent.
Air dry the cuff. DO NOT iron the cuff cover.
Device care
Clean the device with a soft, dry cloth.
Disposal
Batteries and electronic instruments must
be disposed of in accordance with the locally
applicable regulations, and not as domestic
waste.