Appearance of the irregular heartbeat (IHB) symbol
This symbol
BL
indicates that an irregular heartbeat was detected.
In this case, the measured blood pressure may deviate from your
actual blood pressure values. It is recommended to repeat the
measurement.
5. Data memory
This device automatically stores up to 99 measurement values for
each of the 2 users.
Select either user 1 or 2 by pressing the user button
5
.
Viewing the average of the last 28 days
Press the M-button
3
again. The display first shows
«M»
BP
and
«28A»
, which stands for the average measurement values of the
last 28 days.
Blood pressure readings with suboptimal cuff fit
BT
-A are
not considered in the average value.
Viewing the clinical blood pressure average «MyBP»
Pressing the M-button
3
briefly when the device is switched off,
allows you to see the average clinically relevant blood pressure
»MyBP«. The display first shows
«M»
BP
and »MyBP«
BQ
. This
average is only shown when 12 clinically relevant measurement
values in the last 28 days have been performed.
Only measurements that were performed in the morning
between 05:00-10:59 or in the evening between 17:00-
22:59 are considered.
A maximum of 4 readings per day are considered (2 from
the morning and 2 from the evening).
Measurements performed in standard mode and in MAM-
mode are considered in the average, when they have been
performed during the right time of the day.
Measurements performed in MAM-mode or single standard
mode are both classed as single measurements for working
out the »MyBP average«.
Blood pressure readings with suboptimal cuff fit
BT
-A are
not considered in the average value.
Viewing the stored single values
Pressing the M-button again, allows you to see the last performed
measurement. The display first shows
«M»
BP
and a value, e.g.
»M17«. This means that there are 17 single values in the memory.
Pressing the M-button again displays the previous value. Pressing
the M-button repeatedly enables you to move from one stored
value to another.
Pay attention that the maximum memory capacity of 99
memories is not exceeded.
When the 99 memory is full,
the oldest value is automatically overwritten with the
100 value.
Values should be evaluated by a doctor before
the memory capacity is reached – otherwise data will be
lost.
Clearing all values
Make sure the correct user is activated.
If you are sure that you want to permanently remove all stored
values, hold down the M-button (the device must have been
switched off beforehand) until
«CL ALL»
appears and then
release the button. To permanently clear the memory, press the
Bluetooth/Time button while
«CL ALL»
is flashing.
Individual
values cannot be cleared.
Cancel deletion:
press ON/OFF button
1
while
«CL
ALL»
is flashing.
6. Battery indicator and battery change
Low battery
When the batteries are approximately ¾ empty the battery symbol
AS
will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery - replacement
When the batteries are flat, the battery symbol
AS
will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment
AK
at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in Section
»2. Using the device for the first time«.
The memory retains all values although date and time must
be reset – the year number therefore flashes automatically
after the batteries are replaced.
Information for the doctor in case of repeated appearance of
the IHB symbol:
This device is an oscillometric blood pressure monitor that also
measures the pulse during blood pressure measurement and
indicates when the heart rate is irregular.