8. Error Messages

If an error occurs during the measurement, the measurement is 

interrupted and an error message, e.g. «

Err 3

», is displayed.

* Please immediately consult your doctor, if this or any other 

problem occurs repeatedly.

9. Safety, Care, Accuracy Test and Disposal

Safety and protection

Follow instructions for use. This document provides important 

product operation and safety information regarding this device. 

Please read this document thoroughly before using the device 

and keep for future reference.

This device may only be used for the purposes described in 

these instructions. The manufacturer cannot be held liable for 

damage caused by incorrect application.

This device comprises sensitive components and must be treated 

with caution. Observe the storage and operating conditions 

described in the «Technical Specifications» section.

Protect it from:

- water and moisture

- extreme temperatures

- impact and dropping

- contamination and dust

- direct sunlight

- heat and cold

The cuffs are sensitive and must be handled with care.

Do not exchange or use any other kind of cuff or cuff connector 

for measuring with this device.

Only pump up the cuff once fitted.

Do not use this device close to strong electromagnetic fields such 

as mobile telephones or radio installations. Keep a minimum 

distance of 3.3 m from such devices when using this device.

Do not use this device if you think it is damaged or notice 

anything unusual.

Never open this device.

If the device is not going to be used for a prolonged period the 

batteries should be removed.

Read the additional safety information provided within the indi-

vidual sections of this instruction manual.

If you suffer from a cardiac arrhythmia consult with your doctor 

before using the device. See also chapter «Appearance of the 

Irregular Heartbeat (IHB) Symbol» of this user manual.

The measurement results given by this device is not a diag-

nosis. It is not replacing the need for the consultation of a physi-

cian, especially if not matching the patient’s symptoms. Do not 

rely on the measurement result only, always consider other 

potentially occurring symptoms and the patient’s feedback. 

Calling a doctor or an ambulance is advised if needed.

Error

Description Potential cause and remedy

«

Err 1

»

AS

Signal too 

weak

The pulse signals on the cuff are 

too weak. Re-position the cuff and 

repeat the measurement.*

«

Err 2

»

BT

-B

Error signal During the measurement, error signals were 

detected by the cuff, caused for instance by 

movement or muscle tension. Repeat the 

measurement, keeping your arm still.

«

Err 3

»

BT

-C

Abnormal 

cuff 

pressure

An adequate pressure cannot be gener-

ated in the cuff. A leak may have occurred. 

Check that the cuff is correctly connected 

and is not too loose. Replace the batteries 

if necessary. Repeat the measurement.

«

Err 5

» Abnormal 

result

The measuring signals are inaccurate and 

no result can therefore be displayed. Read 

through the checklist for performing reli-

able measurements and then repeat the 

measurement.*

«

Err 6

» MAM Mode  There were too many errors during the 

measurement in MAM mode, making it 

impossible to obtain a final result. Read 

through the checklist for performing reli-

able measurements and then repeat the 

measurement.*

«

HI

»

Pulse or cuff 

pressure too 

high

The pressure in the cuff is too high (over 

299 mmHg) OR the pulse is too high (over 

200 beats per minute). Relax for 5 minutes 

and repeat the measurement.*

«

LO

»

Pulse too 

low

The pulse is too low (less than 40 beats 

per minute). Repeat the measurement.*

Problem 

with Blue-
tooth

®

 

connection

If any problem occurs with the Bluetooth 
connection, the Bluetooth

® 

icon 

AL

 

blinks rapidly for approximately 10 

seconds. To solve the problem, please 

visit www.microlife.com/connect.

Содержание BPB3 BT

Страница 1: ...yhnite sa tesnému alebo priliehavému odevu na hornej časti ramena AIzogibajte se tesnim oblačilom Izbegavajte usku odeću na nadlaktici Kerülje a vastag vagy szoros ruhákat a felkaron Izbjegavajte usku odjeću na nadlaktici Place the artery mark on the cuff over your artery Umieść mankiecie na tętnicy Поместите манжету так чтобы значок артерии на манжете находился над артерией руки в сгибе локтя Пос...

Страница 2: ...n Vyhněte se jídlu koupání kouření nebo kofeinu přibližně 30 minut Vyhnite sa jedlu kúpaniu fajčeniu alebo kofeínu približne 30 minút Izogibajte se hrani kopanju in kajenju približno 30 minut Izbegavajte hranu kupanje pušenje i kofein približno 30 minuta Kerülje az evést fürdést dohányzást vagy koffeinfogyasztást a megelőző kb 30 percben Izbjegavajte jesti kupati se pušiti i kofein oko 30 min Avoi...

Страница 3: ...nfor mation on our products Stay healthy Microlife AG This device uses the same measuring technology as the award winning BP 3BTO A model tested according to the British and Irish Hypertension Society BIHS protocol Table of Contents 1 Using the Device for the First Time Inserting the batteries Setting the date and time Selecting the correct cuff Selecting the user Selecting standard or MAM mode 2 ...

Страница 4: ...t the minutes and pressed the MAM button the date and time are set and the time is displayed 5 If you want to change the date and time press and hold the MAM button for approx 3 seconds until the year number starts to flash Now you can enter the new values as described above Selecting the correct cuff Microlife offers different cuff sizes Select the cuff size to match the circumference of your upp...

Страница 5: ...The cuff will now pump up automatically Relax do not move and do not tense your arm muscles until the measurement result is displayed Breathe normally and do not talk 4 The cuff fit check AR on the display indicates that the cuff is perfectly placed If the icon BT A appears the cuff is fitted suboptimally but it is still ok to measure 5 When the correct pressure is reached the pumping stops and th...

Страница 6: ...elect either user 1 or 2 by pressing the user button 5 Viewing the average of the last 28 days Press the M button 3 briefly when the device is switched off The display first shows M BP and 28A which stands for the average of measurement values of the last 28 days Blood pressure readings with suboptimal cuff fit BT A are not considered in the average value Viewing the clinical blood pressure averag...

Страница 7: ...alues although date and time must be reset the year number therefore flashes automatically after the batteries are replaced Which batteries and which procedure Use 4 new long life 1 5 V size AA alkaline batteries Do not use batteries beyond their date of expiry Remove batteries if the device is not going to be used for a prolonged period Using rechargeable batteries You can also operate this devic...

Страница 8: ...e See also chapter Appearance of the Irregular Heartbeat IHB Symbol of this user manual The measurement results given by this device is not a diag nosis It is not replacing the need for the consultation of a physi cian especially if not matching the patient s symptoms Do not rely on the measurement result only always consider other potentially occurring symptoms and the patient s feedback Calling ...

Страница 9: ...over in a tumble dryer WARNING Under no circumstances may you wash the inner bladder Accuracy test We recommend this device is tested for accuracy every 2 years or after mechanical impact e g being dropped Please contact your local Microlife Service to arrange the test see foreword Disposal Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations n...

Страница 10: ...s reserved The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Micro life Corp is under license Other trademarks and trade names are those of their respective owners Expected service life Device 5 years or 10000 measurements Accessories 2 years ...

Отзывы: