4. Taking a blood pressure measurement

1. Press the ON/OFF button 

1

 to start the measurement.

2. The cuff will now pump up automatically. Support your arm in a 

relaxed position with the device at the same height as your heart. 

Relax, do not move and do not tense your arm muscles until the 

measurement result is displayed. Breathe normally and do not 

talk.

Support your arm in a relaxed position and ensure that the 

device 

is at the same height as your heart.

3. The cuff fit check 

AR

 on the display indicates that the cuff is 

perfectly placed. If the icon 

AR

-A appears, the cuff is fitted subop

-

timally, but it is still ok to measure.

4. When the correct pressure is reached, the pumping stops and the 

pressure falls gradually. If the required pressure was not reached, 

the device will automatically pump some more air into the cuff.

5. During the measurement, the pulse indicator 

AO

 flashes in the 

display.

6. The result, comprising the systolic 

AL

 and the diastolic 

AM

 blood 

pressure and the pulse rate 

AN

 are displayed. Note also the expla

-

nations on further display symbols in this booklet.

7. When the device has finished measuring, remove the cuff.

8. Switch off the device. (The monitor does switch off automatically 

after approx. 1 min.).

You can stop the measurement at any time by pressing the 

ON/OFF button (e.g. if you feel uneasy or an unpleasant pres

-

sure sensation). 

How not to store a reading

As soon as the reading is displayed press and hold the ON/OFF 

button 

1

 until 

«M»

 

AP

 is flashing. Confirm to delete the reading by 

pressing the time button 

4

.

«CL»

 is displayed when the reading is deleted from the 

memory successfully.

How do I evaluate my blood pressure

The triangle on the left-hand edge of the display 

AT

 points at the 

range within which the measured blood pressure value lies. The 

value is either within the optimum (white), elevated (hatched gray) or 

high (black) range. The classification corresponds to the following 

ranges defined by international guidelines (ESH, AHA, JSH). Data in 

mmHg.

Table for classifying blood pressure values in adults in accordance 

with the World Health Organisation (WHO) in 2003. Data in mmHg.

Range

Systolic Diastolic Recommendation 

1 blood pressure 

too high

≥135

≥85

Seek medical 

advice

Содержание BP W10

Страница 1: ...10 EN Master 2022 Revision Date 2022 04 26 BP Wrist Blood Pressure Monitor EN 1 W10 Microlife AG Espenstrasse 139 9443 Widnau Switzerland www microlife com Microlife UAB P Luk io g 32 08222 Vilnius Li...

Страница 2: ...Microlife BP W10 AP AK AL AM AN AT 9 8 AR AO AQ A B C 5 3 1 4 2 6 BP W10 Guarantee Card Name of Purchaser Serial Number Date of Purchase Specialist Dealer...

Страница 3: ...light indicator AK Date Time AL Systolic value AM Diastolic value AN Pulse rate AO Pulse indicator AP Stored value AQ Battery display AR Cuff fit check A Suboptimal cuff fit B Arm movement indicator...

Страница 4: ...ollaboration with physicians and clinical tests prove its measurement accuracy to be very high Batteries and electronic devices must be disposed of in accordance with the locally applicable regulation...

Страница 5: ...nt facts about blood pressure 2 Using the device for the first time Inserting the batteries Setting the date and time 3 Checklist for taking a reliable measurement 4 Taking a blood pressure measuremen...

Страница 6: ...ke at least two readings every time in the morning before taking medications and eating in the evening before going to bed bathing or taking medication and average the measurements It is quite normal...

Страница 7: ...utton for approx 3 seconds in the setting mode Press time button again and then the year number starts to flash Now you can enter the new values as described above 3 Checklist for taking a reliable me...

Страница 8: ...isplayed Note also the expla nations on further display symbols in this booklet 7 When the device has finished measuring remove the cuff 8 Switch off the device The monitor does switch off automatical...

Страница 9: ...readings with suboptimal cuff fit AR A are not considered in the average value Pay attention that the maximum memory capacity of 30 memories is not exceeded When the 30 memory is full the oldest valu...

Страница 10: ...vice for the first time The measurements stored in the memory are deleted when the batteries are removed from the battery compartment e g when replacing batteries Which batteries and which procedure U...

Страница 11: ...n Potential cause and remedy ERR 1 Signal too weak The pulse signals on the cuff aretoo weak Re position the cuff and repeat the measurement ERR 2 Error signal During the measurement error signals wer...

Страница 12: ...infant or neonates Presence of significant cardiac arrhythmia during measurement may interfere with blood pressure measurement and affect the reliability of blood pressure readings Consult with your d...

Страница 13: ...y or for an extended period of time Prolonged exposure to cuff pressure will reduce peripheral perfu sion and may lead to injury Avoid situations of extended cuff pres surization beyond normal measure...

Страница 14: ...on and the safety of usage Protect the device and accessories from the following to avoid damaging the device water other liquids and moisture extreme temperatures impacts and vibrations direct sunlig...

Страница 15: ...arantee The following items are excluded from the guarantee Transport costs and risks of transport Damage caused by incorrect application or non compliance with the instructions for use Damage caused...

Страница 16: ...f method Phase I systolic Phase V diastolic Measurement range SYS 60 200 mmHg DIA 40 200 mmHG 40 199 beats per minute pulse Cuff pressure display range 0 299 mmHg Resolution 1 mmHg Static accuracy wit...

Отзывы: