Make sure that the cuff is positioned 2 cm above the elbow.

The 

artery mark

 on the cuff (ca. 3 cm long bar) must lie over 

the artery which runs down the inner side of the arm.

Support your arm so it is relaxed.

Ensure that the cuff is at the same height as your heart.

6. Any improper setup or damage to the stethoscope will cause 

distorted sound or poor sound transmission causing inaccurate 

readings.

7.

Proper deflation rate is essential for an accurate reading.

 

Practice and master a recommended deflation rate of 2-3 mmHg 

per second or a drop of 1-2 marks on the manometer 

1

 for each 

heartbeat.

Measuring procedure

1. Place the chest piece 

6

 underneath the cuff 

2

 or 1-2 cm 

below it. Make sure the chest piece is in contact with the skin 

and lies on the brachial artery.

2. Plug in the ear piece 

7

 and check, if the chest piece is placed 

correctly, so that the Korotkoff sounds appear the loudest.

3. Close the valve 

4

 on the pump ball 

3

 by turning the screw 

clockwise. Do not over-tighten.

4. Take the pump ball 

3

 in your free hand (the arm you are not 

using to measure) and pump up the cuff. Watch the pressure indi-

cator on the manometer 

1

 and pump up to approx. 40 mmHg 

higher than the expected systolic value (the upper value).

Inflate to 200 mmHg if you are not sure about the expected 

value.

5. Open the valve 

4

 slowly by turning the screw counter clock-

wise whilst holding the stethoscope chest piece 

6

 on the 

brachial artery. Listen carefully as the cuff begins to deflate. 

Note the reading on the manometer 

1

 as soon as you hear a 

faint, rhythmic tapping or thumping sounds. 

This is the 

systolic blood pressure reading.

6. Allow the pressure to continue dropping at the same deflation 

rate. Note the reading on the manometer 

1

 as soon as the 

thumping sound stops. 

This is the diastolic blood pressure 

reading.

7. Deflate the cuff completely.

8. Repeat the measurement at least two further times and record 

your values, date and time immediately after finishing the 

measurements.

9. Remove the cuff and the stethoscope.

4. Malfunction / Troubleshooting

If problems occur when using the device, the following points 

should be checked and if necessary, the corresponding measures 

are to be taken:

If you think the results are unusual, please read through the 

information in «Section 1.» carefully.

Description

Potential cause and remedy

The sound transmis-

sion is poor, distorted 

or there is extraneous 

noise.

Check the ear pieces if they are dirty 

or cracked. If not, make sure you 

wear them properly.

Check the tube if it is broken or 

twisted.

Check the chest piece if there is any 

damage.

Make sure the chest piece is in 

contact with the skin and lies on the 

brachial artery. Clean or replace any 

defective parts if found to avoid inac-

curate reading.

The pressure does not 

rise although the pump 

ball 

3

 is pumping.

Make sure that the valve is closed.

Make sure the cuff is properly 

connected to bulb and manometer.

Check if the cuff, tube and/or bulb is 

leaking. Replace the defective parts 

if any.

The deflation rate can 

not be set to 2-

3 mmHg/sec. by 

adjusting the valve 

4

.

Disassemble the valve from pump ball 

to check if there is any blockage in the 

airway of the valve. Clean the blockage 

and try again. If it still does not work, 

replace it to avoid inaccurate readings.

The manometer needle 

is not at 0 ± 3 mmHg at 

rest.

Make sure that the valve is 

completely open for zero check.

If still more than 3 mmHg deviation, 

contact your dealer to recalibrate the 

manometer.

Содержание BP AG1-40

Страница 1: ...AG1 40 EN PL 0219 EN 1 PL 5 MicrolifeAG Espenstrasse139 9443Widnau Switzerland www microlife com CHDE Polska S A Biesiadna 7 35 304 Rzesz w Tel 0 17 229 37 89 Fax 0 17 230 21 14 E mail biuro rzeszow...

Страница 2: ...erie Serijski broj Serijski broj Num r de serie V robn slo V robn slo Sorozatsz m Seri numaras Date of Purchase Date d achat Data d acquisto Datum kupovine Datum kupovine Data cump r rii Datum n kupu...

Страница 3: ...n your country Alternatively visit the internet at www microlife com where you will find a wealth of invaluable infor mation on our products Stay healthy Microlife AG Table of Contents 1 Important Fac...

Страница 4: ...least 15 seconds between two measurements If you are pregnant you should monitor your blood pressure regularly as it can change drastically during this time How do I evaluate my blood pressure The hig...

Страница 5: ...ing 6 Allow the pressure to continue dropping at the same deflation rate Note the reading on the manometer 1 as soon as the thumping sound stops This is the diastolic blood pressure reading 7 Deflate...

Страница 6: ...ine or dishwasher Accuracy test We recommend this device is tested for accuracy every 2 years or after mechanical impact e g being dropped Please contact your local Microlife Service to arrange the te...

Страница 7: ...y Zapraszamy tak e na nasz stron internetow www microlife pl na kt rej mo na znale wiele u ytecznych informacji na temat naszych produkt w Zadbaj o swoje zdrowie Microlife AG Spis tre ci 1 Wa ne zagad...

Страница 8: ...Wielokrotne powt rzenie pomiaru daje bardziej rzetelne rezultaty ni pojedynczy pomiar Zr b przynajmniej 15 sekundow przerw mi dzy kolejnymi pomiarami W czasie ci y nale y regularnie monitorowa ci nie...

Страница 9: ...na odczyt na manometrze 1 aby jak najszyb ciej us ysze cichy rytmiczny stukanie lub dudni cy d wi k Jest to ci nienie skurczowe ci nienie krwi 6 Pozostaw do dalszego spadku ci nienia na tym samym pozi...

Страница 10: ...lgotnej szmatki oraz mydlin UWAGA Nie pra mankietu w pralce ani zmywarce Sprawdzanie dok adno ci Zaleca si sprawdzenie dok adno ci pomiarowej urz dzenia co 2 lata lub zawsze gdy poddane zostanie ono w...

Страница 11: ...9 BP AG1 40 PL...

Отзывы: