RFT9739 Rack-Mount Transmitter Instruction Manual
83
Appendix
A
RFT9739 Specifications
Performance specifications
Sensor model
Mass flow accuracy*
ELITE
liquid
±0.10% ± [(zero stability / flow rate) x 100]% of rate
gas
±0.50% ± [(zero stability / flow rate) x 100]% of rate
F-Series
liquid
±0.20% ± [(zero stability / flow rate) x 100]% of rate
gas
±0.70% ± [(zero stability / flow rate) x 100]% of rate
D (except DH38), DT and DL
liquid
±0.15% ± [(zero stability / flow rate) x 100]% of rate
gas
±0.65% ± [(zero stability / flow rate) x 100]% of rate
DH38
liquid
±0.15% ± [(zero stability / flow rate) x 100]% of rate
gas
±0.50% ± [(zero stability / flow rate) x 100]% of rate
Sensor model
ELITE
liquid
±0.05% ± [½(zero stability / flow rate) x 100]% of rate
gas
±0.25% ± [(zero stability / flow rate) x 100]% of rate
F-Series
liquid
±0.10% ± [½(zero stability / flow rate) x 100]% of rate
gas
±0.35% ± [(zero stability / flow rate) x 100]% of rate
D (except DH38), DT and DL
liquid
±0.05% ± [½(zero stability / flow rate) x 100]% of rate
gas
±0.30% ± [(zero stability / flow rate) x 100]% of rate
DH38
liquid
±0.05% ± [½(zero stability / flow rate) x 100]% of rate
gas
±0.25% ± [(zero stability / flow rate) x 100]% of rate
Sensor model
Density accuracy
Density repeatability
g/cc
kg/m
3
g/cc
kg/m
3
ELITE (except CMF010P)
liquid only
±0.0005
±0.5
±0.0002
±0.2
ELITE CMF010P
liquid only
±0.002
±2.0
±0.001
±1.0
F-Series
liquid only
±0.002
±2.0
±0.001
±1.0
D6, D12, D25, D40, DH100, DH150
liquid only
±0.002
±2.0
±0.001
±1.0
DH6, DH12, DH38
liquid only
±0.004
±4.0
±0.002
±2.0
D65, DL65, DT65, D100, DT100,
D150, DT150, DH300
liquid only
±0.001
±1.0
±0.0005
±0.5
D300, D600, DL100, DL200
liquid only
±0.0005
±0.5
±0.002
±2.0
Sensor model
Temperature accuracy
Temperature repeatability
All sensors
±1°C ± 0.5% of reading in °C
±0.02°C
* Flow accuracy includes the combined effects of repeatability, linearity, and hysteresis. All specifications for liquids are based on
reference conditions of water at 68 to 77 °F (20 to 25°C) and 15 to 30 psig (1 to 2 bar), unless otherwise noted. For values of zero
stability, refer to product specifications for each sensor.
Содержание Fisher Rosemount RFT9739
Страница 1: ...Model RFT9739 Rack Mount Transmitter Instruction Manual Version 3 Transmitters February 2000 ...
Страница 2: ......
Страница 4: ......
Страница 20: ...10 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 26: ...16 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 92: ...82 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 100: ...90 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 110: ...100 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 112: ...102 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 120: ...110 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 122: ...112 RFT9739 Rack Mount Transmitter Instruction Manual ...
Страница 129: ......