Options
Description
HI
x
> A
The event occurs when the value of the assigned process variable (
x
) is
greater than the setpoint (
Setpoint A
), endpoint not included.
LO
x
< A
The event occurs when the value of the assigned process variable (
x
) is
less than the setpoint (
Setpoint A
), endpoint not included.
IN
A ≤
x
≤ B
The event occurs when the value of the assigned process variable (
x
) is
in
range,
that is, between
Setpoint A
and
Setpoint B
, endpoints included.
OUT
x
≤ A or
x
≥ B
The event occurs when the value of the assigned process variable (
x
) is
out of range,
that is, less than
Setpoint A
or greater than
Setpoint B
, end-
points included.
3.
Assign a process variable to the event.
4.
Set values for the required setpoints.
•
For HI and LO events, set
Setpoint A
.
•
For IN and OUT events, set
Setpoint A
and
Setpoint B
.
5.
(Optional) Configure a discrete output to switch states in response to the event
status.
Related information
6.5
Configure HART/Bell 202 communications
HART/Bell 202 communications parameters support HART communications with the
transmitter's primary mA terminals over a HART/Bell 202 network.
•
Configure basic HART parameters
(Section 6.5.1)
•
Configure HART variables (PV, SV, TV, QV)
(Section 6.5.2)
•
Configure burst communications
(Section 6.5.3)
6.5.1
Configure basic HART parameters
Display
OFF-LINE MAINT
>
OFF-LINE CONFG
>
HART
ProLink III
Device Tools
>
Configuration
>
Meter Information
Device Tools
>
Configuration
>
Communications
>
Communications (HART)
Field Communicator
Configure
>
Manual Setup
>
HART
>
Communications
Integrate the meter with the control system
94
Micro Motion
®
Compact Density Meters (CDM)
Содержание CDM100M
Страница 8: ...Getting Started 2 Micro Motion Compact Density Meters CDM ...
Страница 12: ...Before you begin 6 Micro Motion Compact Density Meters CDM ...
Страница 20: ...Configuration and commissioning 14 Micro Motion Compact Density Meters CDM ...
Страница 26: ...Introduction to configuration and commissioning 20 Micro Motion Compact Density Meters CDM ...
Страница 110: ...Integrate the meter with the control system 104 Micro Motion Compact Density Meters CDM ...
Страница 116: ...Operations maintenance and troubleshooting 110 Micro Motion Compact Density Meters CDM ...
Страница 124: ...Transmitter operation 118 Micro Motion Compact Density Meters CDM ...
Страница 194: ...Using the transmitter display 188 Micro Motion Compact Density Meters CDM ...
Страница 208: ...Using ProLink III with the transmitter 202 Micro Motion Compact Density Meters CDM ...
Страница 218: ...Concentration measurement matrices derived variables and process variables 212 Micro Motion Compact Density Meters CDM ...
Страница 219: ...Concentration measurement matrices derived variables and process variables Configuration and Use Manual 213 ...