Page 55
Software TIM Connect
thermoIMAGER TIM
1
IR image from the camera
2
Temperature profile: Shows the temperatures along max. 2 lines at any size and position in the
image.
3
Reference bar: Shows the scaling of temperature within the color palette.
4
Temperature of measure area: Analyses the temperature according to the selected shape, e.g.
average temperature of the rectangle. The value is shown inside the IR image and the control
displays.
5
Control displays: Displays all temperature values in the defined measure areas like Cold Spots, Hot
Spots, temperature at cursor, internal temperature and chip temperature.
6
Alarm settings: Bar showing the defined temperature thresholds for low alarm value (blue arrow)
and high alarm value (red arrow). The color of numbers within control displays changes to
red
(when temperature above the high alarm value) and to
blue
(when temperature below the low
alarm value).
7
Histogram: Shows the statistic distribution of single temperature values.
8
Automatic / manual scaling of the palette (displayed temperature range): Man., </> (min, max),
1
s
: 1 Sigma, 3
s
: 3 Sigma
9
Icon for quick access to Image Subtraction function.
10
Icon enabling switching between color palettes.
Further details to the software are in the enclosed CD.