Page 49
Electrical Installation
thermoMETER CTL
7.5.1
Entering of the Calibration Code
Every sensor has a specific calibration code, that is clearly printed on a label on the sensor
Fig. 14 Calibration code
i
Please do not remove this label or make sure the code is noted anywhere. The code is needed if the
sensor has to be exchanged.
Every sensor has a specific calibration code, which is printed on the sensor. For a correct temperature mea-
surement and functionality of the sensor this calibration code must be stored into the controller.
The calibration code consists of five blocks with 4 characters each.
Example: EKJ0 -
0OUD -
0A1B
A17U
93OZ
block 1
block 2
block 3
block 4
block 5
i
After exchanging a sensor the calibration code of the new sensor must be entered into the controller.
For entering the code, please press the
and
-Taste (keep pressed) and then the
The display shows
HCODE
and then the 4 signs of the first block. With
and
each sign can be changed.
switches to the next sign or next block.