Page 45
Electrical Installation
thermoMETER CTL
7.3
Cable Assembling
The cable gland M12x1.5 allows the use of cables with a diameter of 3 to 5 mm.
Remove the isolation from the cable (40 mm power supply, 50 mm signal outputs, 60 mm functional
inputs).
Cut the shield down to approximately 5 mm and spread the strands out.
Extract about 4 mm of the wire isolation and tin the wire ends.
Place the pressing screw, the rubber washer and the metal washers of the cable gland one after the
other onto the prepared cable end
.
Spread the strands and fix the shield between two of the metal washers.
Insert the cable into the cable gland until the limit stop.
Screw the cap tight.
Every single wire may be connected to the according screw clamps according to their colors.
Metal washer
Rubber washer
Pressing screw
034J - PQKE - 0JKH
Shield
Fig. 9 Cable Assembling
i
Use shielded cables only.
The sensor shield has to be grounded.