Technical Data
12
-
2.4 Air Purge Collar
The lens must be kept clean at all times from dust, smoke, fumes and other contaminants in order to avoid
reading errors. These effects can be reduced by using an air purge collar. Make sure to use oil-free,
technically clean air only.
The needed amount of air (approx. 20 ... 100 l/ min.) depends on the application and the
installation conditions on-site.
Protective window
Description
Spectral range
Transmission
1)
Germanium
Protective window (50.8 x 3 mm/ Ge) for TIM QVGA-G7, TIM VGA-G7
7.9 µm
0.96
Protective window (50.8 x 3 mm/ Ge) for TIM 160, TIM QVGA, TIM QVGA-HD,
TIM 640
7.5-13 µm
0.92
Zinc sulfide
Protective window (50.8 x 3 mm/ ZnS) for TIM QVGA-G7, TIM VGA-G7
7.9 µm
0.93
Protective window (50.8 x 3 mm/ ZnS) for TIM 160, TIM QVGA, TIM QVGA-HD,
TIM 640
7.5-13 µm
0.91
Protective window (50.8 x 3 mm/ ZnS) for TIM 200, TIM 230 (appropriate for VIS
+ IR)
7.5-13 µm
0.91
Protective window (50.8 x 3 mm/ ZnS) for LT models of thermoMETER CTL, CSL
serie
7.5-13 µm
0.91
Borofloat
Protective window (50.8 x 3 mm/ Borofloat 33) for TIM M-1, TIM M-05
0.85-1.1 µm, 500-540 nm
0.92
1) The displayed values are standard values and can vary between different delivery units.
Содержание Outdoor protective housing
Страница 1: ...Operating Instructions Outdoor Protective Housing for thermoIMAGER TIM ...
Страница 9: ... Technical Data 9 2 2 Dimensions Figure 1 Dimensions with wall mount dimensions in mm not to scale ...
Страница 10: ... Technical Data 10 Figure 2 Dimensions for wall mount dimensions in mm not to scale ...
Страница 16: ... Installation 16 The USB server is now ready mounted Figure 9 Figure 9 USB server ready mounted ...
Страница 18: ... Installation 18 Step 2 Carefully remove the mounting plate Figure 11 Figure 11 Remove the mounting plate ...
Страница 33: ......