Page 34
Assembly
optoNCDT 1900 / Profinet
5.4.6
Plug-In Connection, Supply and Output Cable
optoNCDT
0.3 m
ILD1900-x-IE with pigtail
Do not bend the sensor cable more tightly than 30 mm (fixed installation) or 75 mm
(permanently flexible).
i
The firmly connected sensor cable is drag-chain suitable.
i
Unused open cable ends must be insulated to protect against short circuits or sensor malfunctions.
MICRO-EPSILON recommends the use of the PC1900-IE drag-chain compatible standard connection cables from the optional acces-
sories see Chap.
.
Fasten the plug connection of the cable plug and socket when using a drag-chain compatible PC1900-IE sensor cable.
Avoid excessive pull on the cables. If a cable of over 5 m in length is used and it hangs vertically without being secured, make
sure that some form of strain relief is provided close to the plug connection.
Do not twist a mated connection.
Connect the cable shield to the potential equalization (PE, protective earth conductor) on the evaluator (control cabinet, PC hous-
ing) and avoid ground loops.
Never lay signal lines next to or together with power cables or pulse-loaded cables (e.g., for drives or solenoid valves) in a single
bundle or duct. Always use separate ducts.