Page 54
Set Sensor Parameter
optoNCDT 1220
7.5.3.4
Calculation of Measuring Value using Analog Current
Current output (without zeroing, without teaching)
Variables
Value range
Formula
I
OUT
= current [mA]
[3,8; <4] SMR reserve
[4; 20] measuring range
[>20; 20,2] EMR reserve
d
[mm]
=
* MR
[mm]
(I
OUT
[mA]
- 4)
16
MR = measuring range [mm]
{10/25/50/100/200/500}
d = distance [mm]
[-0,01MR; 1,01MR]
Current output (with zeroing), reference value is midrange
Variables
Value range
Formula
I
OUT
= current [mA]
[3,8; <4] SMR reserve
[4; 20] measuring range
[>20; 20,2] EMR reserve
d
[mm]
=
* MR
[mm]
(I
OUT
[mA]
- 12)
16
MR = measuring range [mm]
{10/25/50/100/200/500}
ZP = zero position [mm]
[0; MR]
Current output (with teaching)
Variables
Value range
Formula
I
OUT
= current [mA]
[3,8; <4] SMR reserve
[4; 20] measuring range
[>20; 20,2] EMR reserve
d
[mm]
=
16
*|n
[mm]
- m
[mm]
|
(I
OUT
[mA]
- 4)
MR = measuring range [mm]
{10/25/50/100/200/500}
m, n = teachig area [mm]
[0; MR]
d
= distance [mm]
[m; n]