Page 100
Operation
colorSENSOR CFO
7
The Auto-Reset is only useful for a triggered measurement, otherwise the last value will remain until a
new trigger is set. Via the
Reset
function the output is reset to output
Color not recognized
in
between.
8
In the input field
Hold time
you can enter a time in ms. If the color group is detected, the output
configuration (10) for this color group is held for the selected time.
During this time the outputs are not updated and thus, further colors are not detected.
i
Production errors occurring in the meantime are not detected, also see chapter Hold time
.
9
Here, the learned color values, depending on the selected color space, are displayed and can also be
edited. The color space can be changed in the detection profile
10
The output configuration defines the outputs to be switched upon recognition of this color group.
The outputs can be set to
high
= blue or
low
= white by clicking on the output configuration.
i
The color table is automatically sorted in the output configuration sequence.
11
Every color is described by the color space coordinates and the permissible tolerance values. Up to
three tolerance limits per color can be specified depending on the distance model (12). If the current
measured value is within these tolerances, the color is recognized and indicated
If the controller detects a color within the tolerance limits, it indicates the status on the left side of the
window with a depiction of the current color.