Page 86
Appendix | Control Menu
C-Box/2A
A 3.2
Tab Settings
A 3.2.1
Inputs
Sensor 1 /
Sensor 2
Sensor status /
Sensor type /
Serial number
Selection of the connected sensor. It supports sensors of the ILD2300, ILD1420, ILD1750,
ILD1900, IFC2421 and IFC2422 series. Additional information
Averaging
No averaging
Measurement values are not averaged.
Moving average Mit-
telwert over
N values
2/ 4 / 8 / 16 / 32 / 64 / 128
The arithmetic mean value Mgl is generated
and output via the selectable filter width N of
consecutive measured values
Recursive average
over N values
Value
Each new metric MW is weighted to the (n-1)
value of the previous average
Median filter over
N values
3 / 5 / 7 / 9
The Median is generated from a preselected
filter width N of measured values
Laser
On / Off
Turns on or off the laser light source on the sensor.
Digital input
Function
Disabled
The multifunction input has no function.
Master C-Box/2A
value
Multifunction input is master pulse input for the C-Box/2A.
i
For this function to work mastering must be enabled
Forward to sensor 1
Multifunction input is forwarded to the corresponding input of
the connected sensor 1.
Forward to sensor 2
Multifunction input is forwarded to the corresponding input of
the connected sensor 2.
Forward to sensor 1
and 2
Multifunction input is forwarded to the corresponding inputs of
the connected sensors 1 and 2.
Logic for digital input
Low-level logic
Settings, see also Trigger Mode chapter
or Synchro-
High-level logic
Содержание C-Box/2A