Page 44
Setting Controller Parameters
C-Box/2A
6.5.4
Synchronization
Synchronization
Synchronization
No synchronization
Sync off. The measuring rate can be freely
adjusted. Value range: from 0.4 to 80 kHz.
Internal synchronization
The C-Box/2A forms the time base.
External
synchroniza-
tion
Low-level logic (LLL)
≤ 0.7 V: Trigger not active
≥ 2.2 V: Trigger active
The Sychronization signal
is received from an ex-
ternal signal source, e.g.
function generator.
High-level logic (HLL)
≤ 3.0 V: Trigger not active
≥ 8.0 V: Trigger active
i
External synchronization is not possible when edge or level triggering is active.
All sensors can be synchronised from the C-Box/2A. A synchronization between them of sensors of the same
type is then no longer necessary. Sensors with different measuring ranges from the same series can be syn-
chronized.
The C-Box/2A operates as Master; the sensors operate as Slave. The small time offset of the measured value
between individual sensors no longer applies. The controller only reacts to the edge of a synchronization
signal.
6.5.5
Output Data Rate
Output data rate
Output every ... measured value Value
The reduction of the output rate causes
only every nth measured value to be
output. The other measured values are
discarded. Any desired averaging over n
values must be set separately.
Reducing applies for following
interfaces
Analog
Ethernet
USB
Fields with a gray
background re-
quire a selection.
Value
Fields with dark
border require the
specification of a
value.
Содержание C-Box/2A
Страница 95: ......