SF82 Transmitter User’s Manual
5
Michell Instruments
If longer cable runs are required, off the shelf 5-pin M12 cables can be connected between the
SF82 transmitter and the cable provided by Michell Instruments.
Note: The cable screen should only be connected to ground point at either the
transmitter installation side or at the receiving equipment. Failure to observe this
precaution can result in ground loops and equipment malfunction.
2.3.
Cable Connection
DIN 43650 Version
To ensure the specified ingress protection is achieved, when installing the connector, the securing
screw (with the O-ring and washer) must be tightened to a minimum torque of 3.4 Nm (2.5 ft-
lbs). The sensor cable used must be a minimum diameter of 4.6 mm (0.2”).
M12 5-Pin Version
The connector should be installed by aligning the locating pin on the transmitter with the slot
on the cable. The connector can then be pushed into place and rotated until finger tight.
2.4.
Electrical Schematic
Note: The cable screen should only be connected to a ground point at either the
transmitter installation side, or at the receiving equipment. Failure to observe this
precaution can result in ground loops and equipment malfunction.
Figure 5 Connector Installation
Figure 6 2-Wire Connection Diagram
O-ring and
washer
Содержание SF82 Series
Страница 1: ...SF82 Dew Point Transmitter User s Manual 97576 Issue 1 June 2019 nbn Austria GmbH...
Страница 23: ...SF82 Transmitter User s Manual 97576 Issue 1 June 2019 16 APPENDIX A Technical Specifications...
Страница 26: ...SF82 Transmitter User s Manual 19 Michell Instruments Product Dimensions M12 5 8 UNF M12 G1 2 M12 3 4 UNF...
Страница 27: ...SF82 Transmitter User s Manual 97576 Issue 1 June 2019 20 MiniDIN 5 8 UNF MiniDIN G1 2 MiniDIN 3 4 UNF...
Страница 28: ...SF82 Transmitter User s Manual 21 Michell Instruments 5 8 UNF 3 4 UNF G1 2 Quick Connect...
Страница 29: ...SF82 Transmitter User s Manual 97576 Issue 1 June 2019 22 APPENDIX B Quality Recycling Warranty Information...
Страница 31: ...SF82 Transmitter User s Manual 97576 Issue 1 June 2019 24 APPENDIX C Return Document Contamination Declaration...
Страница 32: ...SF82 Transmitter User s Manual 25 Michell Instruments Return Document Decontamination Declaration...
Страница 33: ...SF82 Transmitter User s Manual 97576 Issue 1 June 2019 26 APPENDIX D Modbus Register Map...