OPERATION
3 OPERATION
3.1 Preparation
3.1.1 Installing Relative Humidity Instruments for Calibration
Relative humidity probes can easily be installed into the humidity chamber through
the ports in the door. The amount and size of the ports are supplied per customer
speci
fi
cation.
Port openings
available per
customer specification
Blind stop
Figure 13
Probe Ports
NOTE: Ports which are not being used should be covered with a blind stop
to exclude temperature and relative humidity infl uences from outside the
chamber.
NOTE: If a door without ports (A000268) is used, the probes to be tested
must be placed inside the measurement chamber. To ensure good air fl ow do
not overfi ll the chamber with probes.
The port adapters can be changed using a port adapter removal tool (A000265) available
from Michell Instruments.
Figure 14
Port Adapter Removal Tool
Insert the two pegs on the adapter tool into the corresponding holes on the port
adapter and turn counter-clockwise to loosen and clockwise to tighten.
Figure 15
Port Adapters and Removal Tool