Michell Instruments Optidew Vision Скачать руководство пользователя страница 2

OPTIDEW VISION, Panel Mount Fitting  

 

Users Guide Supplement No. 1 

 

 

 

Page 2 

 

 

 

 

www.michell.com 

Introduction. 

 
This guide is for the use of customers who wish to panel mount a standard Optidew 
Vision  Optical  Dewpointmeter  into  a  standard  19”  racking  system.  To  do  this 
properly, a Michell Instruments Ltd. panel mount kit Part No.9980504 is required.  
 

Warning! 

 

Before  starting  work,  ensure  that  the  mains  lead  is  fully  removed  from  the  unit. 
Remove all connections to the instrument (including the sensor and temperature 
probe).

  

 

Dismantling 

 
1. 

Remove the 4 screws holding the rear panel in place (FIG 1). The rear panel will 
now be loose and held by its cables. Take care not to damage the wires. 

2. 

Remove the 2 front panel plastic clip retainers and remove the 4 screws holding 
the front panel (FIG 2). The display PCB should now be loose – allow it to drop 
forward and unplug connectors PL1, 2 & 3. 

3. 

On the underside, remove the 4 pozi drive screws (the nearest to the side panels)  
to release the base plate (FIG 3). 

4. 

Retain all the screws and plastic clips. 

5. 

With the feet resting on a surface, carefully lift the case lid up, by its handle, from 
the  base  plate  until  some  resistance  is  felt.  Then  slide  the  lid  towards  the  front 
until some resistance is felt. 

6. 

Feed the front panel through the bezel and remove the cover. 

7. 

Remove the handle and its fixings from the case lid. Retain the screws. 

8. 

Remove the case feet (1 screw each) and discard. 

9. 

Fit  the  2  small  blocks  to  the  inside  of  the  case  lid  using  the  M3  x  10  long 
countersink head screws (x2) supplied. The screws should be fitted in the middle 
holes  of  the  blocks.  Ensure  that  the  outer  holes  are  in  alignment  with  the 
respective holes in the case lid. 

10. 

Re-assemble  the  instrument  without  the  handle  and  feet.  Ensure  that  the 
connectors  PL1,  PL2  &  PL3  are  re-connected  correctly  and  that  there  are  no 
trapped wires when assembling the cover (SEE FIG 4). 

11. 

Fit the 2 panel mounting brackets to the panel using the M4 x 8 pozi countersink 
head screws (x4) supplied. 

12. 

Slide  the  instrument  through  the  panel  aperture  and  screw  the  case  to  the 
mounting brackets using the M3 x 16 long pan head screws, M3 shakeproof and 
M3 plain washers (x4 of each) supplied. 