MDM300 I.S. Sensor & Battery Replacement Manual
4
97469 Issue 1, October 2014
MAINTENANCE
1.2.4 Battery Replacement
Battery Replacement
To replace the battery, proceed as follows:
1. Open the instrument case as described in Section 1.2.1.
2. Loosen the 3 battery clamp retaining screws (2).
3. Release the battery (1) at connector (3) and remove.
4. Re-connect new battery to connector (3).
5. Replace and tighten battery clamp retaining screws (2).
6. Close the casing as described in Section 1.2.2.
7. Switch on the instrument and check that it is operating satisfactorily.
8. Charge the battery, when necessary, as described in the User’s
Manual.
NOTE: No other parts of the MDM300 I.S. are user-servicable - please contact
your Michell representative if further help is needed.