APPENDIX C

Appendix C 

Recommended Practices in Humidity Measurements

The following text is reproduced with kind permission from the National Physical Laboratory. It is originally published in 

the booklet, 

A Guide to the Measurement of Humidity.

Definition of Relative Humidity

Relative Humidity

 – The ratio of the actual vapor pressure to the saturation vapor pressure over a plane 

liquid water surface at the same temperature, expressed as a percentage. This is commonly understood when 

the term ‘X percent relative humidity’ is used. 

For actual vapor pressure, e, and saturation vapor pressure, e

s

                                          

 e

       

relative humidity (in %) =  –––   x  100

                                             

 e

s

USAGE: 

The  phrase  ‘relative  humidity’  is  commonly  abbreviated  RH  although  this  is  not  a  recognized 

abbreviation. Values of relative humidity are commonly expressed in units of percent relative humidity (% RH).

Recommended practices in humidity measurements

 

General practical recommendations

• 

Where relative humidity is of interest, a direct measurement of relative humidity is usually 

best. Where an absolute measure of humidity is needed, choose dew point, vapor pressure 

or similar measurements.

• 

Establish  the  measurement  requirements  at  the  purchasing  stage  in  order  to  have  the 

right instrument for the job.

• 

Allow hygrometers to equilibrate in any new environment. This is particularly necessary 

after changes in temperature due to transportation or storage. Depending on the instrument 

and on how great the change in conditions, this may require from only a few minutes to 

many hours.

• 

Follow Michell Instruments’ care instructions for the instrument. Some instruments need 

routine cleaning or other maintenance. Before using any solvent cleaner, check with Michell 

Instruments that this will not harm the sensor or other materials of construction.

• 

Wherever possible, ensure that hygrometers are calibrated under the conditions of use, 

i.e. at similar values of humidity and temperature, and (if relevant) in similar conditions of 

pressure, airflow, etc.

• 

Keep a record of calibrations and any adjustments to the hygrometer. This will show the 

long-term stability of the instrument and allow the associated uncertainty to be assessed.

• 

Check  instruments,  if  possible,  at  intervals  between  calibrations,  by  comparison  with 

another (stable) instrument, to monitor for long-term drift. Routine checks are also useful 

before and after subjecting an instrument to transportation or other stress, which might 

lead to a shift in its performance. Where the check is against two (or more) instruments 

this is even better: not only does this add confidence, but in the event of one instrument 

drifting among a set of three, it can be seen which reading is most suspect.

• 

Cleanliness of the environment will affect different hygrometers in different ways. Dust 

and airborne droplets should be avoided or filtered out if possible. Contaminants can come 

from the most surprising sources, ordinary urban pollution, for example.

• 

The  readings  given  by  some  types  of  hygrometer  are  sensitive  to  gas  type.  For  any 

Instrument which reads in terms of mass per unit volume, e.g. in grams per cubic metre, 

it must be confirmed whether the calibration is valid for the gas in use.

• 

Avoid using instruments in direct sunlight or near any other source of heat, unless they 

are suitably shielded to prevent measurement errors.

Содержание HygroSmart I7000XP

Страница 1: ...HygroSmart I7000XP Interchangeable Sensor for RH Temperature User s Manual 97481 Issue 1 January 2015...

Страница 2: ...this information when contacting Michell Instruments for service purposes Sensor Code Serial Number Invoice Date Location of Instrument Tag No Sensor Code Serial Number Invoice Date Location of Instru...

Страница 3: ...copied or otherwise reproduced communicated in any way to third parties nor stored in any Data Processing System without the express written authorization of Michell Instruments Ltd HygroSmart I7000X...

Страница 4: ...re 1 Operating Graph 2 Figure 2 Electrical Connections 3 Figure 3 HygroSmart I7000XP Sensor Pin Connector 4 Figure 4 Hygrosmart I7000XP Operating Conditions 7 Figure 5 Hygrosmart I7000XP Dimensions 7...

Страница 5: ...is designed to be completely safe when used with options and accessories supplied by the manufacturer for use with the instrument Toxic Materials The use of hazardous materials in the construction of...

Страница 6: ...ty s second T temperature V volt Warnings The following general warning listed below is applicable to this instrument It is repeated in the text in the appropriate locations Where this hazard warning...

Страница 7: ...ive humidity and temperature in a wide spectrum of industrial and heavy industrial applications The sensor incorporates the latest HygroSmart 3 technology providing excellent performance across a wide...

Страница 8: ...ensor must be made bearing in mind the effectiveness of air circulation The point at which the sensor is installed must be typical of the surrounding environment where the measurements of relative hum...

Страница 9: ...t I7000XP Pin 1 Not Connected 2 Power Supply 3 Power Supply ve 4 Output Voltage RH 5 Output Temperature PT100 6 Output Temperature PT100 7 Output Temperature PT100 PT100 5 1 2 3 4 6 7 PT1000 PT100 PT1...

Страница 10: ...he protection filter mounted on the measurement probe body NOTE Be careful not to apply any force to the filter protection cage 3 Taking care not to bend the male pins on the sensor electrical connect...

Страница 11: ...HygroSmart I7000XP User s Manual Michell Instruments 5 APPENDIX A Appendix A Technical Specifications...

Страница 12: ...Measurement Range 40 to 85 C 40 to 185 F Temperature Accuracy 0 2 C Temperature Resolution Infinite Electrical Specifications Output Signal White housing 0 1 V Black housing 0 2 1 V Temperature Output...

Страница 13: ...NDIX A A 1 Operating Conditions 25 40 0 50 25 70 85 75 50 100 RH Temp C Maximum mixing ratio 250g water kg dry air Figure 4 Hygrosmart I7000XP Operating Conditions A 2 Dimensions 17 5mm 0 69 12mm 0 47...

Страница 14: ...HygroSmart I7000XP User s Manual 8 97481 Issue 1 January 2015 APPENDIX B Appendix B Quality Recycling Warranty Information...

Страница 15: ...which may be harmful to the environment Similarly we are increasingly using recyclable and or recycled material in our business and products wherever it is practical to do so To protect natural resour...

Страница 16: ...EU Directive 2011 65 EU Article 3 24 defines Control and Monitoring Equipment specifically as monitoring and control instruments designed exclusively for industrial or professional use RoHS2 EU Direc...

Страница 17: ...intains a hazardous material register in which MSDS data sheets are collated and we will check that our suppliers will comply to REACH requirements for all materials and substances we use in the proce...

Страница 18: ...g courses of action will be taken If the fault is covered under the terms of the warranty the instrument will be repaired at no cost to the owner and returned If Michell Instruments determines that th...

Страница 19: ...HygroSmart I7000XP User s Manual Michell Instruments 13 APPENDIX C Appendix C Recommended Practices in Humidity Measurements...

Страница 20: ...ichell Instruments care instructions for the instrument Some instruments need routine cleaning or other maintenance Before using any solvent cleaner check with Michell Instruments that this will not h...

Страница 21: ...of the air between the droplets Such results may exceed 100 RH or may be impossible to interpret meaningfully Droplets of liquid also damage some electrical types of humidity sensor Filtering the air...

Страница 22: ...rials At low humidity anything much below a dew point of 0 C the amounts of water given off by organic and porous materials can dramatically affect the value of humidity The lower the level of moistur...

Страница 23: ...d Filters and saturation guarding may protect the sensor but these actions carry a risk of damage by condensation or other contamination Protective filters can slow the response time of sensors This c...

Страница 24: ...HygroSmart I7000XP User s Manual 18 97481 Issue 1 January 2015 APPENDIX D Appendix D Return Document Decontamination Declaration...

Страница 25: ...cals YES NO Radioactive substances YES NO Other hazards YES NO Please provide details of any hazardous materials used with this equipment as indicated above use continuation sheet if necessary Your me...

Страница 26: ...http www michell com...

Отзывы: