Easidew Portable HP User’s Manual
Michell Instruments
13
MAINTENANCE
•
Replace the locking screw and cable gland.
5.3
Filter Cartridge Replacement
Unscrew the gas inlet
fi
tting to reveal the
fi
lter, which can then simply be removed for
checking or replacement.
5.4
Checking the Easidew Portable Electronics Calibration
h
h
Supply
4-20 mA
PIN 3
PIN 1
(GND)
Screen/Shield
Optional
Figure 3
2-Wire Connection - View Showing Rear of Connector Terminal Block
To verify the display electronics are within calibration, a 4-20mA current source can be
connected in place of the sensor.
At 4 mA the display should show -100.0°Cdp ±0.5°C (-148°Fdp ±0.9°F)
At 20 mA the display should show +20.0°Cdp ±0.5°C ( +68°Fdp ±0.9°F)
If the displayed values are outside of this range, please contact Michell Instruments'
Service Department.
5.5
List of Spares
P/N
Description
SSF-PF-10PK
Pack of 10 particulate
fi
lter cartridges
EAP-BAT
Replacement battery pack
EAP-CHR
Replacement battery charger