Cermet II User’s Manual
Michell Instruments
27
APPENDIX A
Appendix A
Technical Speci
fi
cations
Monitor
Display
5 digit LED
7-segment digit, 14.2 mm(0.56”)
Display Color
Red
Temperature and Alarm
Indicators
Red LED’s
Performance Speci
fi
cations
Measurement Ranges
-19999 to 99999
Accuracy
Dew point ±0.5°C (±1.0°F)
ppm
V
, lbs/MMSCF and g/m
3
±1% of reading
Pressure ±0.5% full-scale reading
Electrical Speci
fi
cations
Dew-Point Input Channel
Digital signal from Michell dew-point transmitter
Pressure Input Channel
(optional)
2-wire pressure transducer, 4-20 mA (loop-powered)
Output Scaled Range
Dew point: -100 to +20°C
Moisture content in gas: 0 to 9999 ppm
V
, 0 to 1000 lbs/MMSCF
and g/m
3
Non-standard available upon request
Re-transmittion Output
Signals
Standard
: 4-20 mA
Optional
: 0-20 mA (max load 500
Ω
) or 0 - 10 V optional (min
load 5 K
Ω
)
User scaleable
Alarm Relays
Standard
: Two relays SP1 and SP2, fully user programmable
Form C contacts rated 10 A, 240 V AC or 8 A, 24 V DC
Non-inductive load
Optional
: Two extra relays, SP3 and SP4, fully user programmable
Form A contacts rated 5 A, 240 V AC or 5 A, 30 V DC
Non-inductive load
Power Connection
2m (6.56ft), 3 wire
Power Supply
Standard
: 85 to 265 V AC, 50/60 Hz or 95 to 370 V DC
Optional
: 18 to 36 V AC or 9 to 60 V DC
Power Consumption
10 W max
Transmitter Cable
Copper braid screened cable, 4 core 7/0.2 (0.22mm²), stranded,
tinned copper conductors, PVC, insulated, Melinex taped, black
PVC outer
Fitted with transmitter connector and terminations for monitor
Max length 1000m (3280ft)
Operating Speci
fi
cations
Operating Mode
Continous
Operating Temperature
0 to +50°C (+32 to +122°F)
Mechanical Speci
fi
cations
Ingress Protection
Standard
: IP54 / NEMA 12 (at front),
Optional
: Extra protection cover IP66 / NEMA 4
Dimensions
96 x 48 x 142mm (3.8 x 1.9 x 5.6")
Mounting
Panel mounting (1/8 DIN cut out 92 x 45mm (3.62 x 1.77”))
Weight
0.6kg (1.32lbs)
Содержание Cermet II
Страница 33: ...Cermet II User s Manual 26 97049 Issue 25 3 October 2016 APPENDIX A Appendix A Technical Specifications...
Страница 36: ...Cermet II User s Manual Michell Instruments 29 APPENDIX B Appendix B Setup Codes...
Страница 42: ...Cermet II User s Manual Michell Instruments 35 APPENDIX D Appendix D EU Declaration of Conformity...
Страница 43: ...Cermet II User s Manual 36 97049 Issue 25 3 October 2016 APPENDIX D Appendix D EU Declaration of Conformity...
Страница 44: ...Cermet II User s Manual Michell Instruments 37 APPENDIX E Appendix E Quality Recycling Warranty Information...
Страница 51: ...Cermet II User s Manual 44 97049 Issue 25 3 October 2016 NOTES...
Страница 52: ...http www michell com...