The product has calibrated before out of factory. After correctly wiring, it can be
powered and used immediately. But for further precise signal sampling, user needs to
warm-up the product for 3 minutes before use.
a)
The auxiliary power requires isolation voltage
≥
2000V
AC
, DC current output ripple
<
10mV, and multiple converters can share with one set auxiliary power together.
b)
Output signal (Iz) is designed to follow the impedance load value of
250
Ω
;
When
Resistance (R
1
) is less than
100
Ω
,
the user must add output current circuit a power
reduced resistance load R
0
to keep the summation of total impedance value is
between
100
Ω~
250
Ω
3.
Basic testing method for accuracy
(1)
According to the terminal definition table to connect the testing circuit;
(2)
The testing must to be done by under the following conditions
;
Auxiliary power:
24V
±
1
%,
ripple
≤
5mV
Ambient temperature:
25
℃±
2
℃
Relative humidity
: RH(45
~
75)
%
Accuracy for Signal Power Source instrument:
0.05
Three
Phase
Power
Source
Auxiliary Power
1
8
7
6 7
V2
+
-
Transducer
U
A
U
B
3
5
I
C
I
B
I
A
U
C
N
2(4)
250Ω
7
8
Mianyang Weibo Electronics Co.,Limited
No.7 Road 2 of Xianren Road Youxian District Mianyang City P.R. China
Tel: 86-816-2971265 Fax: 86-816-2281934
www.wb-my.com