Note:
If your problem cannot be solved by the solutions
above, contact Customer Service at (877) 358-4342
(Mon–Fri, 9am–5pm, EST). Do not disassemble
the unit.
11
“Err P” displayed
Fail to inflate
Return for servicing.
“Err 1” displayed
Deflate too rapidly Return for servicing.
“Err 2” displayed
Movement during
measurement
Do not move during
measurement.
Signal
interference
Remove interfering source
e.g. mobile phones,
magnets.
“Err 3” displayed
Incorrect results
Measure again.
Problem Cause
Solution
No display when
you press the Power
/Start button
Have the batteries
run out?
Are the battery
polarities wrong?
Replace new batteries.
Re-align batteries in correct
polarities.