Max. respiration rate allowed (triggered hold-off time):

Fixed @ 30 BPM in R/M mode

Higher altitudes allow slightly higher maximum respiration rates (BPM):

0-10K ft. = 20 BPM, 12-20K ft. = 22 BPM, 21 - 25 = 25 BPM and 26+ = 30 BPM

Apnea time-out envelope:

Fixed @ 45 sec.

The apnea alarm does not respond if in any of the "D" modes and while below that pressure altitude threshold

Operating inlet pressures:

1 bar (15 psig.) DYNAMIC (flowing) through cannula and 1.5 meters (5 ft.) of 4 mm. inlet tubing
2 bar (30 psig.) STATIC (non-flowing) 2.2 bar (35 psig) MAX.

Operating Voltage & Current @ 25° C. @ 25% RH. (measured in "N" mode setting @ 15 RESP/ typical.):
Low-v. cut-out:

5.00 VDC+0  mv.  -250 mv.

(red LED on steady)

Min.:

5.50 VDC @ 2.25 ma. idle (100 ma. peak, 3.25 ma. mean)

(Low batt. red LED winking 1/sec.)

Low:

6.50 VDC @ 2.25 ma. idle (100 ma. peak, 3.25 ma. mean)

(Low batt. red LED winking 2/sec.)

Nom.:

7.85 VDC @ 2.25 ma. idle (100 ma. peak, 3.25 ma. mean)

Max.:

11.0 VDC @ 4.30 ma. idle (100 ma. peak, 3.25 ma. mean)

(EPS provides 10 VDC)

Reverse voltage protection:
Diode shunt (1 amp peak max., no internal fuse or over voltage protection provided)
External Power is provided via a 3.5 mm. coaxial power-jack system. Inside (tip) is negative (-) Outside (ring) is positive (+)

Battery Life:
Operating:

40 to 60 Hr's @ 25° C. @ 25% R.H.
Measured from mean, assuming fresh alkaline battery operating under normal operating conditions.

Storage:

Approx. 4-6 months while in the off setting. (battery drain is about 1 µa. in the off setting)
Make sure that the battery is removed from the unit during long-term storage.

Battery type:
Standard 9 volt alkaline DURACELL type MN1604 or equivalent.

Operating temperature and humidity (assumes nominal operating voltage):
Min.:

0% RH @ -55° C to +80° C.

Nom.:

25% RH @ +25° C.

Max.:

100% RH @ +50° C. near condensing

Allowable vibration during operation:
Random vibration 5 to 500 Hz, 15 minutes per axis @ 2.5 g. (rms.) sin wave;

Physical characteristics (EDS-D1 unit only):
Width:

2.60" (6.60 cm.)

Height A:

4.10" (10.41 cm.) enclosure only

Height B:

6.25" (15.87 cm.) enclosure and inlet/outlet tubes with connectors

Depth:

1.50" (3.81 cm.)

Weight:

8.0 Oz. (0.227 kg.) with battery

Physical characteristics Regulator only (XCR type standard issue):
Width A:

1.25" (3.17 cm.)

Width B:

2.10" (5.33 cm.)

Height:

2.25" (5.71 cm.)

Weight:

4.0 Oz. (0.113 kg.)

Page 8 of 8

Height

Width

Depth

Front view

Top view

 100 mm.

Height

Width

A

B

Outlet

CGA-540 or DIN 477 

cylinder inlet

Testing and characterization was done under normal
operating conditions i.e. EDS at 25°C and responding
to a respiration rate of about 16 breaths per minute
without any flow fault or apnea detection's.

General Specifications

This is a list of specification parameters for the EDS model D-1.  These specifications are operational and performance standards or

limits against which the unit was tested or calculated.  Specifications are subject to change without notice.

Содержание EDS-D1

Страница 1: ...evice is intended for supplemental breathing while exposed to high altitudes On demand and up to 25 BPM this device delivers a pulse of supplied oxygen at the start of each inhalation phase Higher alt...

Страница 2: ...n High E S Company assumes no responsibly for any accidents injury or death that may result from the misuse of this device equipment This includes any usage of this device equipment outside the scope...

Страница 3: ...g tests are showing that 90 of the oxygen the EDS D1 provides to the lungs goes directly into the blood The EDS D1 synchronous inhalation pulse demand technique is currently the most efficient way kno...

Страница 4: ...hing in while pushing door outwards fig 1 This is best done if the unit is held with the battery door facing up at you and out The EDS D1 unit uses a standard 9 nine volt alkaline battery connector sy...

Страница 5: ...Positive GRN RED indicators for FLOW NO FLOW oxygen delivery status Provides Apnea alarm to inform user of kinked pinched disconnected lines or holding breath too long Saves existing oxygen supply by...

Страница 6: ...10 000 ft 3 0 2 5 2 0 1 5 1 0 5 0 PRESS ALT X 1K PULSE UNITS D 5 D 1 0 D 1 2 D Modes Unit delays activation until at and above selected pressure altitudes D12 Switch as seen in one of the Day or Delay...

Страница 7: ...r for use with regulators that are flow rate adjustable F MODES Floor Face Mask These next four settings deserve your attention The F Mode settings are F10 F15 F20 and F25 called the floor or facemask...

Страница 8: ...negative and the outside ring is positive A reverse polarity diode 1 A in the EDS unit helps prevent damage from accidental short term reverse voltage occurrences Misc notes The EDS model D1 has a bas...

Страница 9: ...itially off turn on the cylinder valve slowly to allow gas from the regulator to back charge the outlet side of the EDS D1 unit Next turn the EDS D1 unit on During the time the EDS D1 unit initializes...

Страница 10: ...r normal operating conditions Storage Approx 4 6 months while in the off setting battery drain is about 1 a in the off setting Make sure that the battery is removed from the unit during long term stor...

Страница 11: ...r for use with an EDS unit NOTICE IMPORTANT EDS INLET PRESSURE NOTICE If The EDS D1 unit s will not be used with the XCR XCP micro lite system or EDS STR regulator the alternate regulator must be able...

Страница 12: ...s and capability never offered before Complete single and multi place carry on EDS system packages are available along with padded fabric carry strap down harnesses for most of our kevlar fiber and al...

Отзывы: