Advanced cable tester

ENGLISH

ENG

ENG

KPS SOLUCIONES EN ENERGÍA, S.L.

Parque Empresarial de Argame,

C/Picu Castiellu, Parcelas i-1 a i-3

E-33163 Argame, Morcín

Asturias, España, (Spain)

See the full

instruction manual:

WARNING

This cable locator is produced in accordance with the safety specifications for electronic 

meters  and  testing  instruments  and  has  been  completely  tested  before  packing  and 

transport. Before using this device, please read this manual carefully and follow all of the 

instructions. Failure to stand by these instructions or ignoring the warnings and cautions 

herein may lead to personal hurt, threat to life or damage of equipment.

ATTENTION

Please observe the following instructions to ensure safe operation and optimal performance.

Preliminary inspection

Before the first use, please check whether the cable locator can function normally and ensure it 

is not damaged during storage and transport. If there is any damage, please contact the supplier.

WARNING

Whenever  you  use  the  cable  locator,  the  related  safety  specifications  in  electronics 

industry must be abided by.

Use

The following instructions should be followed to avoid electric shock, short circuit or explosion:

•  This cable locator can be directly used for live parts, but do take insulation measures in 

accordance with industrial safety codes to avoid electric shock and hurt.

•  To  avoid  electrical  shock,  the  valid  safety  and  VDE  regulations  regarding  excessive 

contact voltages must receive utmost attention, when working with voltages exceeding 

120V (60V) DC or 50V (25V) rms AC. The values in brackets are valid for limited ranges 

(such as medicine and agriculture).

•  Never try to make contact between both battery, make certain of correct polarity. Batteries with 

reversed polarity can lead to instruments destruction. Furthermore, they may explode or ignite.

WARNING

•  Measurements in dangerous proximity of electrical installations only under the direction 

of a responsible electrician.

•  When the product is used for testing a live line, do ensure the test lead has been taken 

down  from  the  tested  object  before  connecting  or  taking  down  the  test  lead  of  the 

transmitter, and remind the surrounding persons to be well protected.

•  Never  try  to  disassemble  battery  cells!  The  battery  contains  very  strong  base 

chemicals. Danger of causticization! If the battery contents come in contact with skin or 

clothing, rinse immediately with water. If the battery contents come in contact with eyes, 

immediately flush by using pure water and consult a doctor.

•  As  the  connection  of  the  transmitter  with  the  mains  may  generate  circuit  current  of 

milliampere level, in live condition the grounding hole of the transmitter can be only 

connected with a neutral conductor. If transmitter connection is realized from the phase 

towards the protective conductor, the fucntional safety of the protective conductor must 

be tested first, incompliance with DIN VDE 0100. The reason is that when connecting 

the  transmitter  from  phase  towards  ground,  all  parts  being  connected  to  the  earth 

may be live in the event of an error (if the earth resistance does not comply with the 

prescriptions).

•  If the operator’s safety is no longer ensured, the instrument is to be put out of service 

and protected against use. The safety is no longer ensured, if the instrument:

 -

shows obvious damage

 -

does not carry out the desired measurements

 -

has been stored for too long under unfavorable conditions

 -

has been subject to mechanical stress during transport

•  The instrument may only be used under those conditions and for those purposes for 

which it was conceived. When modifying or changing the instrument, the operational 

safety is no longer ensured.

Repair and maintenance

•  If the detector is suspected of malfunctioning, please confirm that the electrical amperage 

of the battery is sufficient and the test lead is not broken.

•  Before sending back the detector for repair, please dismantle the battery and describe 

the phenomenon of malfunction, and then properly pack the apparatus to avoid damaging 

it during transportation. For damages caused during transportation, the company would 

assume no responsibility.

•  There  is  a  fuse  within  the  transmitter.  If  it  is  damages  during  the  guarantee  period,  it 

can only be replaced by technical personnel of the company. If it is damaged after the 

guarantee period, please replace it with a fuse of the same model by yourself. This fuse is 

single metal wire fast-melting type, so do not replace it with a helical metal wire of delayed 

type, otherwise the transmitting power and safety of apparatus cannot be guaranteed.

Checking of the fuse of the transmitter

The fuse of the transmitter can prevent the transmitter from being damaged by overload or  

wrong operations. If the fuse in the transmitter has already melted down, the trasmitter can 

only transmit weak signals. If the self-check of transmitter is passed but the signal transmitted 

is weak, it means that the transmission is sound but the fuse has been melted down. If no 

signal is found during the transmission status self-check of the transmitter, and the voltage of 

battery is normal, it means that the transmitter is broken and should be repaired by special 

technicians.

Specific methods and steps to check the fuse of the transmitter:

•  Break all the measuring circuits of the transmitter

•  Turn on the transmitter and put it in the transmitting status

•  Set the transmitting power of transmitter to level I

•  Connect one end of the test lead to the joint of the transmitter

•  Insert the other end of test lead into the correcting socket of the transmitter

•  Turn on the transmitter to search for the signals from the test lead, and move the probe of 

the receiver towards the test lead

•  If the fuse is not broken, the value displayed on the receiver will be doubled

Cleaning

Use a rag dipped with clean water orneutral detergent to wipe the transmitter, and then use 

a dry rag to wipe it again.

CAUTIONS

Before cleaning work, please make sure that the equipment has been turned off, and all 

the circuits have been cut off.

During cleaning work, please do not use benzene, alcohol, acetone, ether, ketone, thinner 

or gasoline, which could make the equipment deformed or decolored.

After cleaning, use the equipment again when it is completely dry.

Replace the battery

If the battery symbol on the display flickers, and the buzzer gives warning, then the battery 

must be replaced.

Steps to replace the battery (of the transmitter or the receiver) are as follows:

•  Turn off the equipment and cut off all the measuring circuits

•  Unscrew the back of the equipment, and take out the battery cover of the equipment

•  Take out the battery used up

•  Install a new battery according to correct polarity

•  Put back the battery cover and fasten the screws

WARNING

•  When inserting or replacing the battery, please do pay attention to the correct polarity 

of  battery.  If  the  polarity  of  battery  is  incorrect,  the  equipment  will  be  damaged. 

Furthermore, explosion may be caused.

•  Do not connect the two poles of the battery with a conducting wire, and do not throw the 

battery into fire, otherwise there may be risk of explosion.

•  Please  do  not  try  to  dismantle  the  battery!  The  electrolyte  contained  shows  strong 

basicity,  which  has  the  risk  of  causing  corrosion!  If  the  electrolyte  contacts  skin  or 

clothes, use fresh water to wash the relevant parts. If the electrolyte enters the eyes, 

use fresh water to wash the eyes inmediately, and see a doctor as soons as possible.

CAUTIONS

•  Before  replacing  the  battery,  the  equipment  must  be  turned  off,  all  the  connected 

measuring circuits must be cut off, and all the conducting wires for measurement must 

be removed.

•  Only the battery specified in the technical parameters table can be used.

•  If the equipment will not be used in a relatively long time, take out the battery within it. If 

the detector is contaminated due to leakage of battery, mail the equipment back to the 

original factory for cleaning and checking.

•  When disposing the used batteries, please abide by the existing regulations on the 

recovery, reuse and disposal of batteries.

Interval of calibration

In order to ensure the precision of measurement made by the equipment, the equipment 

should be calibrated regularly by the adjusting personnel of the company. The interval of 

calibration recommended is one year. If the equipment is used frequently or the conditions of 

usage are very poor, the interval of calibration should be shortened. If the equipment is rarely 

used, the interval of calibration can be extended to three years.

ENGLISH

APAC: 

MGL GLOBAL SOLUTIONS LTD

[email protected]

Flat 4-1, 4/F, No. 35, Section 3 Minquan 

East Road. Taipei, Taiwan.

Tel: +886 2-2508-0877

EMEA: 

MGL EUMAN S.L.

[email protected]

Parque Empresarial Argame, 33163 

Morcín. Asturias, Spain.

Tel: +34 985-08-18-70

AMERICAS: 

MGL AMERICA, LLC.

[email protected]  

US East Coast:

 2810 Coliseum Centre 

Drive, Ste. 100. Charlotte, North Carolina, 

28217 USA

Tel: +1 833 533-5899

US West Coast:

 760 Challenger Street. 

Brea, California 92821 USA

Tel: +1 310-728-6220

www.mgl-intl.com

Отзывы: