Operation and Maintenance Manual 

Manual 118 Issue 15 

 

 

Page 7 of 28 

3.  INSTRUCTIONS FOR USE 

3.1  Before Operating Unit 

BEFORE operating your new 

SA

 suction unit please read the following instructions carefully. 

Become thoroughly familiar with the operation and maintenance of the unit before using and note the 
information on the control panel and rear of the unit.  Only persons trained in its use should operate 
the unit. 

Check that the voltage supply is as printed on the rating plate fitted to the back of the 

SA

 suction 

unit. 

Unscrew and remove the transport screws located on the underside of the unit.  

Warning!  

Always electrically isolate the 

SA

 suction unit from the mains power supply 

before carrying out any cleaning, maintenance or repairs.  Electrical isolation of 
the 

SA

 suction unit is by removal of the mains power plug from the mains 

power supply. 

3.2  Unit Set Up 

 

Connect the desired collection system securely in place on the bracket fitted to the top of 
the unit.  More detail can be found later in this section. 

 

Connect the desired equipment to the patient inlet tube on the collection container top.  
Multi-fit bubble tubing (If supplied) has an I/D 

7mm. Recommended length of tubing 

used is 2 meters. 

 

Connect the unit to the power supply, check label on rear of unit for correct ratings.  Do 
not obstruct access to the mains power supply, and ensure the power cable does not 
create a trip hazard or is subject to damage.  Ensure the exhaust outlet on the side of the 
unit is kept clear of any restriction.   

 

Switch the 

SA

 suction unit on by pressing the green rocker switch so that it illuminates.  

The lamp indicates the 

SA

 suction unit is operating and ready for use. 

3.3  Handling 

The larger 

SA

 suction unit are not normally intended for lifting or carrying.  They will either be 

designed to installed on a trolley or fitted with lockable wheels.  This mobility allows the larger 

SA

 

suction unit to be moved during use and easily transported between locations. 

3.4  Operating Environment 

Operation of this equipment must be within the following ambient condition 

+40°C

-5°C

%

80%

30%

 

Temperature 

Humidity 

 

Warning!  

Never operate a 

SA

 suction unit in the presence of flammable gas such as 

anaesthetic agents.  This is an Explosion hazard! 

3.5  Gauge (Manometer) - Indicated Vacuum 

The unit has a gauge fitted to allow the vacuum to be set at a pre-determined level and monitored 
during use.  To achieve the desired degree of vacuum, occlude the tubing fitted to the patient inlet on 
the collection container top, and read the indicated vacuum on the gauge.  Adjust the vacuum control 
until the required degree of vacuum is achieved.  (Turning the knob clockwise   increases the 
vacuum).  The 

SA

 suction unit is now set to operate at the selected vacuum. 

Содержание SAM 35

Страница 1: ...Suction Unit Operating and Maintenance Manual M G Electric Colchester Ltd Altbarn Close Colchester Essex CO4 9HY United Kingdom Tel 44 0 1206 842244 Fax 44 0 1206 845849 e mail sales mgelectric co uk...

Страница 2: ...Gauge Manometer Indicated Vacuum 3 6 External Filter 3 7 Internal Filter 3 8 Fluid Collection Containers 3 9 Disposable Liner Systems 3 10 Effect of altitude 3 11 Cleaning Procedure 3 12 Transport 3...

Страница 3: ...dhered to Failure to follow this guideline will invalidate any warranty claims and result in the equipment being destroyed Definition of symbols used in these instructions The Instruction for use must...

Страница 4: ...ersons must be familiar with the relevant standards rules accident prevention regulations and operating conditions as a result of their training experience and instruction They are qualified to carry...

Страница 5: ...a trip hazard or is subject to damage The suction pumps may only be opened by qualified technical personnel Electric shock hazard Equipotential bonding may be required in some critical treatment areas...

Страница 6: ...units are not intended for field and transport use SA 36 General purpose high vacuum medical suction unit incorporating two collection containers of 2L capacity Other types of collection container ar...

Страница 7: ...bel on rear of unit for correct ratings Do not obstruct access to the mains power supply and ensure the power cable does not create a trip hazard or is subject to damage Ensure the exhaust outlet on t...

Страница 8: ...SA suction unit must have fitted internally a hydrophobic filter capsule to prevent ingress of foreign material into the pump The hydrophobic filter prevents costly pump damage as well as additional...

Страница 9: ...nt dependent upon the level of vacuum in the container at the time when the valve closes Warning The overflow valve may not operate fully against frothing To reduce frothing an anti foam agent may be...

Страница 10: ...iner may be sterilised by autoclaving at 121 C or by washing with water based disinfectants Before autoclaving rinse the collection container well to remove any detergent 3 9 2 Abbott Receptal Disposa...

Страница 11: ...ude Altitude effects all vacuum pumps and it should be noted that there will be a reduction in the maximum achievable vacuum negative pressure level equivalent to approximately 3 5 per 300m approximat...

Страница 12: ...equipment is required by a patient for home use Medical Staff must fully instruct the patient on the safe operation of the equipment In the event of equipment contamination or failure the patient mus...

Страница 13: ...hing and gloves MUST be worn to avoid contamination Disposal of contaminated parts must be according to local protocols Warning Any dismantling and re assembly of this equipment for whatever purpose m...

Страница 14: ...icate the internal pump requires maintenance Warning Always electrically isolate the SA suction unit from the mains power supply before carrying out any cleaning maintenance or repairs Electrical isol...

Страница 15: ...SA 35 and SA 36 Operation and Maintenance Manual Manual 118 Issue 15 Page 15 of 28 4 4 General Layout...

Страница 16: ...alance if necessary Remove both pump head nozzles from the pump head Remove the transfer valves Clean the valve seats and check the valves for signs of wear Replace any worn or damaged components as n...

Страница 17: ...ically isolate the SA suction unit from the mains power supply before carrying out any cleaning maintenance or repairs Electrical isolation of the SA suction unit is by removal of the mains power cabl...

Страница 18: ...SA 35 and SA 36 Operation and Maintenance Manual Manual 118 Issue 15 Page 18 of 28 4 8 Flow Diagram SA 35 and SA 36...

Страница 19: ...proved spare parts must be used with all SA suction units failure to use original spares will invalidate the warranty and may cause injury and or damage to unit All spares can be purchased by the user...

Страница 20: ...and fixings 10pk MSP1017 2amp 20mm fuses 10pk MSP1225 Power cable 2M MSP1567 5 2 3 Pump Components MQ45 pump Pump motor 230V 50 60Hz MSP1226 Motor 230V 50 60Hz MSP1227 MQ45 pump head connector set MS...

Страница 21: ...ndard supply voltages available 220V 240V 50 60 Hz 1Ph 110V 120V 50 60 Hz 1Ph Capacitor 220V 250V 4uF 110V 120V 12uF Fuse type 5 x 20mm 220V 250V F2 0A 250V 110V 120V F 3 0A 250V Electrical protection...

Страница 22: ...ent special precautions are needed regarding EMC electromagnetic compatibility It is important that the SA suction unit is configured and installed put into service in accordance with the instructions...

Страница 23: ...ed with other equipment If adjacent or stacked use with other equipment is necessary the SA 35 36 and the other equipment should be observed monitored to verify normal operation in the configuration i...

Страница 24: ...intended installation location to assure that it is sufficiently low NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacturer s declaration electromagnetic im...

Страница 25: ...hs should be less than 3 V m Recommended separation distances between Portable and mobile RF communications equipment and the SA 35 36 The SA 35 36 is intended for use in an electromagnetic environmen...

Страница 26: ...SA 35 and SA 36 Operation and Maintenance Manual Manual 118 Issue 15 Page 26 of 28 7 NOTES...

Страница 27: ...SA 35 and SA 36 Operation and Maintenance Manual Manual 118 Issue 15 Page 27 of 28...

Страница 28: ...or with direct probe SA 52 Low vacuum pipeline regulator with remote probe SA 53 Low vacuum pipeline regulator with direct probe SA 54 High vacuum pipeline regulator remote probe mobile trolley Resear...

Отзывы: