background image

MFJ-836 SWR/Wattmeter and RF Ammeter 

 

Instruction Manual

 

RF Ammeter Calibration

 

 

1.

 

Refer to 

Figure 1

 for the Test Setup and refer to the PCB layout in 

Figure 3 

for trim pot location. 

 

2.

 

Remove the top of the MFJ-836. 

 

3.

 

Connect the Test Setup equipment as shown in 

Figure 1

.  Use a 50-ohm 

dummy load for the antenna.  Set the Transmitter to the 14 MHz in the 20-
meter band. 

 

4.

 

Set the Current Range

 

selector to 1A.  For meaningful results, 

your wattmeter 

must be accurate and the 50-ohm load free of significant reactance.

  To test, 

apply a 50-watt CW signal to the 50-ohm dummy load.  You should observe a 
full-scale reading of 1A (100 on the scale).  If results vary significantly from 
this reading, adjust trim pot R3 for the correct indication.  Trim pot R3 is 
located on the pc board below the sense coil. 

 

 
 
 
 
 

 

 

 

 
 
 
 
 
 

Figure 3

: RF Ammeter PCB Layout and Trim Pot Location 

 
 

IMPORTANT WARNING:  Avoid  direct  physical  contact  with  this  device 

when making measurements with power sources of over 
5 Watts.  RF burns or permanent injury may result from 
contact with high RF voltages. 

 
If your meter is out of calibration and you lack the equipment or skills to perform the 
calibration procedure, you may return it to the MFJ factory for service. 

 

 
 

 
 

 

 

R3 

Содержание MFJ-836

Страница 1: ...1312B in order to function A 9 VDC may also be installed inside the meter with a battery clip that is included The rear panel accepts a 2 1 mm coaxial plug with the center conductor positive and the sleeve negative The negative lead is grounded inside the meter There is no need to disconnect the battery when using external power The RF Ammeter does not require power to operate 3 An internal lamp b...

Страница 2: ...ith different voice characteristics With most voices the PEP reading is three to five times higher than the average voice power reading The most accurate PEP readings are obtained only with sustained carrier voice or two tone test modulation During normal voice modulation the wattmeter will typically indicate only 70 of the true Peak Envelope Power 4 Forward power is displayed on the left hand FOR...

Страница 3: ... to estimate readings quickly 10 To prevent meter pegging begin checks on a high range and work down To estimate unknown current levels use the formula below I Unknown Current in Amps P Applied Power in Watts and R Load Resistance in Ohms I P R Range Meter 300 mA 1 A 3 A 10 30 0 1 0 3 20 60 0 2 0 6 30 90 0 3 0 9 40 120 0 4 1 2 50 150 0 5 1 5 60 180 0 6 1 8 70 210 0 7 2 1 80 240 0 8 2 4 90 270 0 9 ...

Страница 4: ... trim pot location 2 Remove the top of the MFJ 836 3 Connect the Test Setup equipment as shown in Figure 1 Use a 50 ohm dummy load for the antenna Set the Transmitter to the 14 MHz in the 20 meter band 4 With the 3000 300 Watts button pushed out for the 300 Watt scale Transmit 100 Watts as indicated on the reference meter Adjust the LO FWD trim pot to set the forward power scale to 100 Watts 5 Wit...

Страница 5: ...ts as indicated on the reference meter Adjust the HI REF trim pot to set the reflected power scale to 100 Watts 9 SWR requires no calibration Figure 1 SWR Wattmeter and RF Ammeter Test Setup Figure 2 SWR Wattmeter PCB Layout and Trim Pot Location Transmitter Transceiver Standard Reference Wattmeter MFJ 836 SWR Wattmeter and RF Ammeter 50 Ohm Dummy Load HI FWD LO REF HI REF LO FWD ...

Страница 6: ...reactance To test apply a 50 watt CW signal to the 50 ohm dummy load You should observe a full scale reading of 1A 100 on the scale If results vary significantly from this reading adjust trim pot R3 for the correct indication Trim pot R3 is located on the pc board below the sense coil Figure 3 RF Ammeter PCB Layout and Trim Pot Location IMPORTANT WARNING Avoid direct physical contact with this dev...

Страница 7: ...RY C109 100 C108 01 D103 IN4007 12V SW 103B LP LP METER LAMP R109 150k R108 50k 2 1 R110 N L 2 1 R111 68k R112 470 C104 001 U101 LM324 3 2 OUT 6 D102 IN34A 12V SW 102B C106 10 R113 IM U102 LM324 3 2 OUT 6 C107 01 R114 22k R120 25k 2 1 FM FG FORWARD METER Q101 2N3904 SW 101A REF R102 150k R101 50k 2 1 R103 N L 2 1 R104 82k R105 470 C101 001 U103 LM324 3 2 OUT 6 D101 IN34A 12V SW 102A C102 10 R106 I...

Страница 8: ... 0549 or the MFJ Factory at 662 323 5869 You will be best helped if you have your unit manual and all information on your station handy so you can answer any questions the technicians may ask You can also send questions by mail to MFJ Enterprises Inc 300 Industrial Park Road Starkville MS 39759 by Facsimile FAX to 662 323 6551 or by email to techinfo mfjenterprises com Send a complete description ...

Отзывы: