3 Installation
890 Titrando
■■■■■■■■
21
Balance
Cable
Mettler AE with interface option
011 or 012
6.2125.020 + 6.2125.010
Also from Mettler: ME 42500
hand switch or ME 46278 foot
switch
Ohaus Voyager, Explorer, Analyti-
cal Plus
Cable AS017-09 from Ohaus
Precisa balances with RS-232-C
interface
6.2125.080 + 6.2125.010
Sartorius MP8, MC, LA, Genius,
Cubis
6.2134.060
Shimadzu BX, BW
6.2125.080 + 6.2125.010
Operation with Touch Control
Connect the balance as follows:
1
Plug in the USB plug of the USB/RS-232 adapter at the USB connec-
tor of the 890 Titrando.
2
Connect the RS-232 interface of the USB/RS-232 adapter with the
RS-232 interface of the balance (see table for cable).
3
Switch on the Touch Control.
4
Switch on the balance.
5
Activate the RS-232 interface of the balance if necessary.
6
Configure the RS-232 interface of the USB/RS-232 adapter in the
device manager of the Touch Control (see Touch Control manual).
3.4.5
Connecting a PC keyboard (only for operation with Touch Con-
trol)
The PC keyboard is used as an aid for text and numerical input.
Connect the PC keyboard as follows:
1
Insert the USB plug of the keyboard into one of the USB sockets of
the 890 Titrando.