62
▪▪▪▪▪▪▪
Standardization using the Iris Adapter, without spacers.
Reference Standardization may be performed in this order, or “Use Iris” first, and “Use Tray” second.
Select Diagnostics, Reference
Standard,
Create Reference Standard.
Vision prompts for the
instrument reference. The
XDS MultiVial Analyzer has an
internal instrument ceramic
reference which is positioned
automatically.
When Vision first
communicates with the
instrument (or after
disconnection and re-
connection) this screen is
shown.
Vision must establish the
configuration and have it
verified by the user. This
assures that test data is sent
to the correct location in the
Diagnostic Database.
If the Sample Module Serial
Number field is empty, locate
the sample module serial
number on the side of the
module. Record the serial
number and enter it in this
field.
Click “OK to accept the
instrument identification.
Vision briefly displays a dialog box which indicates that the instrument reference is being placed
into position. With the MultiVial Analyzer, this is automatic. No user action is required.
The status bar indicates scan progress.
Vision requests that the Certified 80%
Reflectance Reference Standard be placed
in the sampling area.