Insert a new module for the Multi Autolab Cabinet
Figure 34 – Adding all the M101 or M204 modules to the Multi Autolab Cabinet
5.1.3
– Assign the position number for each M101 or M204 module
Requirement
Connect the M101 or M204 cell cable to each M101 or M204 module in the Multi
Autolab Cabinet before continuing. Short circuit the RE and CE plugs and the S and WE
plugs for each cell cable, as indicated in Figure 35 before continuing with this section.
31 |