4 Appendix
Function
2 potential-free analog signal inputs per system.
Connector
Dsub connector 9 pin (female)
1
5
6
9
Input voltage range
±
2.5 V
Input amplification
1, 2, 4, 8, 16, 32, 64 (selectable)
Sampling rate
10, 20, 30, 50, 60 measuring pooints/s
Resolution
24 bit (1 LSP = 0.298
µ
V at amplification 1)
Noise
< 20
µ
Vpp
< 3
µ
Vrms (at 0 V input voltage, amplification 1
and sampling rate 10 Hz)
Zero error
±
2.5 mV
Pin assignment
Pin 1,4,5,8,9: GND (ground, connected to
housing)
Pin 2:
Analog 1+
Pin 3:
Analog 1–
Pin 6:
Analog 2+
Pin 7:
Analog 2–
Safety specifications
Construction/testing
according to EN/IEC 61010-1 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions
The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
Emitted interference
Standards met:
- EN/IEC 61326
- EN 55022
- CISPR 22
Immunity to interference
Standards met:
- EN/IEC 61326
- EN/IEC 61000-4-2
- EN/IEC 61000-4-3
- EN/IEC 61000-4-4
- EN/IEC 61000-4-5
- EN/IEC 61000-4-6
- EN/IEC 61000-4-8
- EN/IEC 61000-4-11
- EN/IEC 61000-4-14
- NAMUR