3 Operation
750 Autosampler
38
Summary of the remote control commands
Command
Description
@GTV ###
@LDL #
@VLV LD/INJ
@RNP
@RVN
@RVP
@RDX
@RDY
@RDZ
@BZR $$$
@AUX 1 ON/OFF
@MSG ## "$$$"
@LOC
Go to vial position ### :
0
Rest position at left front
1
…
128
Vial position on sample tray
Fill sample loop:
The transfer needle is lowered to position # (1
…
4).
Switch injection valve:
LD
LOAD (switch injection valve to FILL position)
INJ
INJECT (switch injection valve to INJECT position)
Read needle position:
The position (1
…
4) is put out.
Read vial number:
The position (1
…
128; HOME) is put out.
Read injection valve position:
The position (LOAD, INJECT) is put out.
Read x position:
The x position (0
…
760) is put out.
Read y position:
The y position (0
…
1000) is put out.
Read z position:
The z position (0
…
46) is put out.
Trigger acoustic signal $$$:
CLK
single short acoustic signal
ERR
single long acoustic signal
ALM
triple slow acoustic signal
PNC
repeated slow acoustic signal
ON
continuous signal
OFF
switch off acoustic signal (
PNC
,
ON
)
Set contact closure 1 ("T1") on or off.
Display at position ## (1
…
80; 1
…
20 = 1
st
line; 21
…
40 = 2
nd
line;
41
…
60 = 3
rd
line; 61
…
80 = 4
th
line) on the LCD display the message
"$$$...".
Return to normal operation (LOCAL).
#
= digit 0
…
9;
$
= any character