![Metrohm 700 Dosino Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/metrohm/700-dosino/700-dosino_manual_1779550044.webp)
4.2 Dosing accuracy according to ISO 8655-3
700 Dosino, Instructions for use
38
4.2 Dosing accuracy according to ISO 8655-3
Metrohm Ltd. guarantees the following limits (according to ISO
8655-3, draft) for Dosino burets with glass cylinders:
Max. permissible systematic error e
s
Nominal
cylinder volume
Maximum permissible systematic errors e
s
a)
in mL
rel.
abs.
2
± 0,3 %
± 6 µL
5
± 0,3 %
± 15 µL
10
± 0,2 %
± 20 µL
20
± 0,2 %
± 40 µL
50
± 0,2 %
± 100 µL
a)
Expressed as deviation of the mean of a tenfold measurement
from any given volume in the range of 10 to 100 % of the nominal
volume.
s
s
V
V
e
−
=
or in %
(
)
0
100
V
Vs
V
s
e
−
=
V
: mean
s
V
: selected volume
0
V
: nominal volume
Maximum permissible random error s
Nominal
cylinder volume
Maximum permissible random errors
repeatybility s
in mL
CV
b)
s
c)
2
± 0,1 %
± 2 µL
5
± 0,1 %
± 5 µL
10
± 0,07 %
± 7 µL
20
± 0,07 %
± 14 µL
50
± 0,05 %
± 25 µL
b)
Expressed as coefficient of variation of a tenfold measurement
from any given volume in the range of 10 to 100 % of the nominal
volume.
V
s
CV
100
=
c)
Expressed as repeatability standard deviation of a tenfold mea-
surement from any given volume in the range of 10 to 100 % of
the nominal volume.